Beckman Coulter, Inc.: Beckman Coulter UniCel Dxl 600, Part numbers: A71460, A71461, A30260; Immunoassay Method, Troponin Subunit
Read the official notice at FDA (device) →
Get alerts for Beckman Coulter
The text below is copied from the official FDA (device) record. If anything differs, the official notice is correct.
- Company / brand
- Beckman Coulter, Inc.
- Product
- Beckman Coulter UniCel Dxl 600, Part numbers: A71460, A71461, A30260; Immunoassay Method, Troponin Subunit
- Reason / hazard
- Beckman Coulter has determined that, due to a potential communication issue, the analyzers listed above may not successfully restart following a standard shutdown, even after multiple attempts. If this issue occurs, the instrument will be unable to complete initialization and transition to Ready mode and the instrument will not process patient samples, QCs or calibrators. This causes delayed results.
- Classification
- Class II
- Status
- Ongoing
- Distribution / sold at
- Worldwide Distribution: US (nationwide) and OUS (foreign) to countries of: Albania, Algeria, Argentina, Australia, Austria, Azerbaijan, Bahrain, Bangladesh, Belarus, Belgium, Bosnia and Herzegovina, Botswana, Brazil, Bulgaria, Cabo Verde, Canada, Chile, China, Costa Rica, Croatia, Czechia, Denmark, Dominican Republic, Egypt, El Salvador, Eswatini, Ethiopia, France, Georgia, Germany, Ghana, Gibraltar, Greece, Hong Kong, Hungary, India Indonesia, Ireland, Israel, Italy, Japan, Jordan, Kazakhstan, Kenya, Korea, Republic of, Kosovo, Kuwait, Kyrgyzstan, Lebanon, Libya, Lithuania, Malaysia, Mayotte, Mexico, Moldova, Republic of, Morocco, Myanmar, Nepal, Netherlands, New Zealand, Nigeria, North Macedonia, Oman, Pakistan, Panama, Paraguay, Peru, Philippines, Poland, Portugal, Puerto Rico, Qatar, Romania, Russian Federation, San Marino, Saudi Arabia, Serbia, Singapore, Slovakia, Slovenia, South Africa, Spain, Sweden, Switzerland, Taiwan, Province of China, Thailand, Tunisia, Turkey, Ukraine, United Arab Emirates, United Kingdom of Great Britain and Northern Ireland, Viet Nam, and Zimbabwe.
- Recall initiated
- 2025-11-26
- Code info
- A71460: UDI 15099590369224 A71461: UDI 15099590369248 A30260: UDI 15099590341602
- Product quantity
- 1272 units
- Voluntary mandated
- Voluntary: Firm initiated
- City
- Chaska
- State
- MN
- Country
- United States
- Event id
- 98070
Free daily digest + "My products" watchlist