Beckman Coulter, Inc.: DxU 840m Iris, REF: C76947; DxU 850m Iris, REF: C49513
- Recall status
- Ongoing (still active)
- Severity
- Class II: could cause temporary or reversible health problems What recall classes mean
- Where sold
- Worldwide - US Nationwide distribution including in the states and territory of CA, NE, FL, NY, MA, MD, SC, WI, TX, OH, VA, PA, IL, NH, KS, GA, NC, IA, PR, TN, OR, UT, NJ, MI, CO, NM, MS, LA, RI, OK...
Common questions
- Is Beckman Coulter DxU 840m Iris REF: C76947 recalled?
- Yes. FDA (device) recalled Beckman Coulter DxU 840m Iris REF: C76947 on 2026-10-21. Recall status: Ongoing (still active).
- What is the hazard?
- The rinse-line check valve has a shorter service lifespan than originally expected and requires annual replacement to prevent lamina fluid leakage; however, the check valve replacement was excluded from the annual preventive maintenance (PM) requirements without adequate reliability evaluation. When the check valve...
- Where was it sold?
- Worldwide - US Nationwide distribution including in the states and territory of CA, NE, FL, NY, MA, MD, SC, WI, TX, OH, VA, PA, IL, NH, KS, GA, NC, IA, PR, TN, OR, UT, NJ, MI, CO, NM, MS, LA, RI, OK...
Read the official notice at FDA (device) →
More recalls
- Beckman Coulter, Inc.: ROTOR ASSY-SW41, Reference Number 331362, when the rotors
- Beckman Coulter, Inc.: ROTOR ASSY-SW40, Reference Number 331302, when the rotors
- Beckman Coulter, Inc.: IQ200 SPRINT, REF: 700-3322,700-3322RFB, 700-3325
- Beckman Coulter Inc.: DxC 500 AU Clinical Chemistry Analyzers: 1. AU w/out ISE
- Beckman Coulter Inc.: DxC 500i Clinical Analyzer Modules: 1. DxC 500 AU Module
The text below is copied from the official FDA (device) record. If anything differs, the official notice is correct.
- Company / brand
- Beckman Coulter, Inc.
- Product
- DxU 840m Iris, REF: C76947; DxU 850m Iris, REF: C49513
- Reason / hazard
- The rinse-line check valve has a shorter service lifespan than originally expected and requires annual replacement to prevent lamina fluid leakage; however, the check valve replacement was excluded from the annual preventive maintenance (PM) requirements without adequate reliability evaluation. When the check valve is not replaced annually, it may allow lamina fluid to migrate internally and leak onto the table or floor, potentially resulting in exposure to physical hazards (slip/fall), chemical hazards (lamina fluid exposure), biohazardous exposure, property damage, and delayed patient results.
- What to do (remedy)
- Classification
- Class II
- Status
- Ongoing
- Distribution / sold at
- Worldwide - US Nationwide distribution including in the states and territory of CA, NE, FL, NY, MA, MD, SC, WI, TX, OH, VA, PA, IL, NH, KS, GA, NC, IA, PR, TN, OR, UT, NJ, MI, CO, NM, MS, LA, RI, OK, WA, MO, ME, IN, CT, AZ, AR, MT, MN, ID, DE, AL, WV, DC, KY, NV, WY, ND, HI, AK, SD, VT, GU and the countries of Canada, Mexico, Switzerland, Panama, Ecuador, Japan, Hungary, Italy, Belgium, China, Germany, France, United Kingdom of Great Britain and Northern Ireland, Russian Federation, Singapore, Romania, Argentina, Israel, Spain, Belarus, Guatemala, Colombia, Qatar, Czechia, Reunion, Australia, United Arab Emirates, Poland, Croatia, Brazil, Slovakia, New Zealand, Saudi Arabia, Thailand, Republic of Korea, Serbia, Austria, Macao, Ukraine, India, Netherlands, Taiwan, South Africa, Sint Maarten, Bangladesh, Hong Kong, Viet Nam, Bahrain, Uganda, Lebanon, Morocco, Guadeloupe, Greece, Curacao, Republic of Moldova, Malaysia, Ghana, Indonesia, Lithuania, Portugal, Kazakhstan, Georgia, Philippines, Bosnia and Herzegovina, Libya, Kuwait, Oman, Jordan, Uzbekistan.
- Recall initiated
- 2026-08-12
- Code info
- Product/UDI-DI: DxU 840m/15099590748654, DxU 850m/15099590748647. All Serial Numbers
- Product quantity
- 2,088
- Voluntary mandated
- Voluntary: Firm initiated
- City
- Miami
- State
- FL
- Country
- United States
- Event id
- 99832