Beckman Coulter, Inc.: ROTOR ASSY-SW40, Reference Number 331302, when the rotors are installed in the following centrifuges: Optima XPN-100 (Model Numbers A99846 a
- Recall status
- Ongoing (still active)
- Severity
- Class II: could cause temporary or reversible health problems What recall classes mean
- Where sold
- US Nationwide distribution.
Common questions
- Is Beckman Coulter ROTOR ASSY-SW40 Reference Number 331302 when the rotors are recalled?
- Yes. FDA (device) recalled Beckman Coulter ROTOR ASSY-SW40 Reference Number 331302 when the rotors are on 2026-10-28. Recall status: Ongoing (still active).
- What is the hazard?
- Optima XE XPN ultracentrifuge rotors may contain a microstructure defect that can cause rotors to break into two or more pieces (bifurcate) during a centrifugation run, resulting in destruction of samples being processed. When used for in vitro diagnostic (IVD) sample preparation, rotor breakage may delay patient...
- Where was it sold?
- US Nationwide distribution.
Read the official notice at FDA (device) →
More recalls
- Beckman Coulter, Inc.: ROTOR ASSY-SW41, Reference Number 331362, when the rotors
- Beckman Coulter, Inc.: IQ200 SPRINT, REF: 700-3322,700-3322RFB, 700-3325
- Beckman Coulter, Inc.: DxU 840m Iris, REF: C76947; DxU 850m Iris, REF: C49513
- Beckman Coulter Inc.: DxC 500 AU Clinical Chemistry Analyzers: 1. AU w/out ISE
- Beckman Coulter Inc.: DxC 500i Clinical Analyzer Modules: 1. DxC 500 AU Module
The text below is copied from the official FDA (device) record. If anything differs, the official notice is correct.
- Company / brand
- Beckman Coulter, Inc.
- Product
- ROTOR ASSY-SW40, Reference Number 331302, when the rotors are installed in the following centrifuges: Optima XPN-100 (Model Numbers A99846 and A94469), Optima XPN-90 (Model Numbers A99842 and A94468), Optima XPN-80 (Model Numbers A99839 and A95765), Optima XE-100 (Model Numbers A99836 and A94516), and Optima XE-90 (Model Numbers A99833 and A94471)
- Reason / hazard
- Optima XE XPN ultracentrifuge rotors may contain a microstructure defect that can cause rotors to break into two or more pieces (bifurcate) during a centrifugation run, resulting in destruction of samples being processed. When used for in vitro diagnostic (IVD) sample preparation, rotor breakage may delay patient test results. Additionally, sample destruction may expose laboratory personnel to biohazardous materials.
- What to do (remedy)
- Classification
- Class II
- Status
- Ongoing
- Distribution / sold at
- US Nationwide distribution.
- Recall initiated
- 2026-09-21
- Code info
- Serial Numbers: 24U15260 to 26U15764; Optima UDI-DI: A99846 - 15099590416324, A99842 - 15099590415754, A99839 - 15099590415822, A99836 - 15099590415846, A99833 - 15099590415860
- Product quantity
- 92 units
- Voluntary mandated
- Voluntary: Firm initiated
- City
- Indianapolis
- State
- IN
- Country
- United States
- Event id
- 99882