Beckman Coulter, Inc.: ROTOR ASSY-SW41, Reference Number 331362, when the rotors are installed in the following centrifuges: Optima XPN-100 (Model Numbers A99846 a
- Recall status
- Ongoing (still active)
- Severity
- Class II: could cause temporary or reversible health problems What recall classes mean
- Where sold
- US Nationwide distribution.
Common questions
- Is Beckman Coulter ROTOR ASSY-SW41 Reference Number 331362 when the rotors are recalled?
- Yes. FDA (device) recalled Beckman Coulter ROTOR ASSY-SW41 Reference Number 331362 when the rotors are on 2026-10-28. Recall status: Ongoing (still active).
- What is the hazard?
- Rotors have been found with a non-conforming microstructure, which may result in some rotors breaking into two or more pieces during use. Rotors that break into multiple pieces during use may result in significant damage to materials (i.e., samples) being processed in the centrifuge and to the centrifuge instrument.
- Where was it sold?
- US Nationwide distribution.
Read the official notice at FDA (device) →
More recalls
- Beckman Coulter, Inc.: ROTOR ASSY-SW40, Reference Number 331302, when the rotors
- Beckman Coulter, Inc.: IQ200 SPRINT, REF: 700-3322,700-3322RFB, 700-3325
- Beckman Coulter, Inc.: DxU 840m Iris, REF: C76947; DxU 850m Iris, REF: C49513
- Beckman Coulter Inc.: DxC 500 AU Clinical Chemistry Analyzers: 1. AU w/out ISE
- Beckman Coulter Inc.: DxC 500i Clinical Analyzer Modules: 1. DxC 500 AU Module
The text below is copied from the official FDA (device) record. If anything differs, the official notice is correct.
- Company / brand
- Beckman Coulter, Inc.
- Product
- ROTOR ASSY-SW41, Reference Number 331362, when the rotors are installed in the following centrifuges: Optima XPN-100 (Model Numbers A99846 and A94469), Optima XPN-90 (Model Numbers A99842 and A94468), Optima XPN-80 (Model Numbers A99839 and A95765), Optima XE-100 (Model Numbers A99836 and A94516), and Optima XE-90 (Model Numbers A99833 and A94471)
- Reason / hazard
- Rotors have been found with a non-conforming microstructure, which may result in some rotors breaking into two or more pieces during use. Rotors that break into multiple pieces during use may result in significant damage to materials (i.e., samples) being processed in the centrifuge and to the centrifuge instrument.
- What to do (remedy)
- Classification
- Class II
- Status
- Ongoing
- Distribution / sold at
- US Nationwide distribution.
- Recall initiated
- 2026-09-21
- Code info
- Serial Numbers: 24U15250 to 26U15766; Optima UDI-DI: A99846 - 15099590416324, A99842 - 15099590415754, A99839 - 15099590415822, A99836 - 15099590415846, A99833 - 15099590415860
- Product quantity
- 336 units
- Voluntary mandated
- Voluntary: Firm initiated
- City
- Indianapolis
- State
- IN
- Country
- United States
- Event id
- 99882