⚠ RecallRadar📷 ScanGuidesRecentOfficial U.S. recall notices, restated plainly

Beckman Coulter, Inc.: ROTOR ASSY-SW41, Reference Number 331362, when the rotors are installed in the following centrifuges: Optima XPN-100 (Model Numbers A99846 a

Medical DevicesSource: FDA (device) · Date: 2026-10-28

Recall status
Ongoing (still active)
Severity
Class II: could cause temporary or reversible health problems What recall classes mean
Where sold
US Nationwide distribution.

Common questions

Is Beckman Coulter ROTOR ASSY-SW41 Reference Number 331362 when the rotors are recalled?
Yes. FDA (device) recalled Beckman Coulter ROTOR ASSY-SW41 Reference Number 331362 when the rotors are on 2026-10-28. Recall status: Ongoing (still active).
What is the hazard?
Rotors have been found with a non-conforming microstructure, which may result in some rotors breaking into two or more pieces during use. Rotors that break into multiple pieces during use may result in significant damage to materials (i.e., samples) being processed in the centrifuge and to the centrifuge instrument.
Where was it sold?
US Nationwide distribution.

Read the official notice at FDA (device) →

Share on XFacebookRedditEmail

Is Beckman Coulter recalled?

Guides: Which agency handles which recall (FDA vs USDA vs CPSC vs NHTSA)

More recalls

The text below is copied from the official FDA (device) record. If anything differs, the official notice is correct.

Company / brand
Beckman Coulter, Inc.
Product
ROTOR ASSY-SW41, Reference Number 331362, when the rotors are installed in the following centrifuges: Optima XPN-100 (Model Numbers A99846 and A94469), Optima XPN-90 (Model Numbers A99842 and A94468), Optima XPN-80 (Model Numbers A99839 and A95765), Optima XE-100 (Model Numbers A99836 and A94516), and Optima XE-90 (Model Numbers A99833 and A94471)
Reason / hazard
Rotors have been found with a non-conforming microstructure, which may result in some rotors breaking into two or more pieces during use. Rotors that break into multiple pieces during use may result in significant damage to materials (i.e., samples) being processed in the centrifuge and to the centrifuge instrument.
What to do (remedy)

How to claim your remedy

Classification
Class II
Status
Ongoing
Distribution / sold at
US Nationwide distribution.
Recall initiated
2026-09-21
Code info
Serial Numbers: 24U15250 to 26U15766; Optima UDI-DI: A99846 - 15099590416324, A99842 - 15099590415754, A99839 - 15099590415822, A99836 - 15099590415846, A99833 - 15099590415860
Product quantity
336 units
Voluntary mandated
Voluntary: Firm initiated
City
Indianapolis
State
IN
Country
United States
Event id
99882
Free daily digest + "My products" watchlist

Get the daily digest, and tell us what to watch. Everything is optional except your email.

Free allergen alerts:

Categories:

Privacy: we store only what you enter here, use it only to send recall alerts you asked for, never sell or share it, and you can unsubscribe anytime. Double opt-in: we will email you a confirmation link before sending anything; alerts start only after you confirm. Every email has unsubscribe and manage-watchlist links.