Beckman Coulter, Inc.: Access 2 Reaction Vessels, individual, disposable, plastic tubes used to process patient, quality control and calibration samples on the Acc
Read the official notice at FDA (device) →
Get alerts for Beckman Coulter
The text below is copied from the official FDA (device) record. If anything differs, the official notice is correct.
- Company / brand
- Beckman Coulter, Inc.
- Product
- Access 2 Reaction Vessels, individual, disposable, plastic tubes used to process patient, quality control and calibration samples on the Access 2 Immunoassay system, Catalog Number 81901
- Reason / hazard
- It has been determined that certain lots of Access 2 Reaction Vessels may contain manufacturing deformities. The bottom (round end) of the Reaction Vessel may have a small protrusion or hair like deformity. This could cause causing instrument errors (e.g., pipetting, rake, wash carousel errors) that could delay patient results.
- Classification
- Class II
- Status
- Ongoing
- Distribution / sold at
- US Nationwide. Global Distribution.
- Recall initiated
- 2025-11-07
- Code info
- UDI-DI: 15099590180249; Lot Numbers: P243943, P243971, P244202 P244344, P244345, P250001, P250002, P250003, P250004 and P250005
- Product quantity
- 38,225 untis
- Voluntary mandated
- Voluntary: Firm initiated
- City
- Chaska
- State
- MN
- Country
- United States
- Event id
- 97932