Howmedica Osteonics Corp.: TRIDENT 0¿ POLY INSERT 28mm ID 620-00-28D Orthopaedic implant components manufactured from UHMWPE (polyethylene)
Read the official notice at FDA (device) →
Get alerts for Howmedica Osteonics
The text below is copied from the official FDA (device) record. If anything differs, the official notice is correct.
- Company / brand
- Howmedica Osteonics Corp.
- Product
- TRIDENT 0¿ POLY INSERT 28mm ID 620-00-28D Orthopaedic implant components manufactured from UHMWPE (polyethylene)
- Reason / hazard
- Affected devices were manufactured with Ultra High Molecular Weight Polyethylene (UHMWPE) raw material aged in excess of five years. Product has the potential for elevated levels of oxidation, which may result in increased wear, leading to hazards such as excessive wear debris, material fragments, fractured device, or disassociated device. Potential harms include revision surgery, pain, and inflammation.
- Classification
- Class II
- Status
- Ongoing
- Distribution / sold at
- Worldwide distribution - US Nationwide and the countries of Canada, LATAM, Europe, Asia Pacific, and South Africa.
- Recall initiated
- 2026-07-30
- Code info
- 620-00-28D UDI-DI 07613327313581 Lots MHANP2 MHD1N0
- Product quantity
- 12,156 US
- Voluntary mandated
- Voluntary: Firm initiated
- City
- Mahwah
- State
- NJ
- Country
- United States
- Event id
- 99558