⚠ RecallRadar📷 ScanGuidesRecentOfficial U.S. recall notices, restated plainly

Howmedica Osteonics Corp.: TS Femoral Prep Tray, Triathlon. Catalog Number: 6543-6-507.

Medical DevicesSource: FDA (device) · Date: 2026-10-21

Recall status
Ongoing (still active)
Severity
Class II: could cause temporary or reversible health problems What recall classes mean
Where sold
These trays are provided by sale or loan. Loans are provided through Distributor. Foreign Consignees will be identified by the countries impacted. 158 trays have been distributed OUS.

Common questions

Is Howmedica Osteonics TS Femoral Prep Tray Triathlon recalled?
Yes. FDA (device) recalled Howmedica Osteonics TS Femoral Prep Tray Triathlon on 2026-10-21. Recall status: Ongoing (still active).
What is the hazard?
Reprocessing instructions provided in LSTPI-B specify a sterilization dry time of 30 minutes for the trays sterilized in blue wrap. When blue wrap is used, the 30-minute dry time may lead to a small amount of residual moisture remaining after sterilization.
Where was it sold?
These trays are provided by sale or loan. Loans are provided through Distributor. Foreign Consignees will be identified by the countries impacted. 158 trays have been distributed OUS.

Read the official notice at FDA (device) →

Share on XFacebookRedditEmail

Is Howmedica Osteonics recalled?

Guides: Which agency handles which recall (FDA vs USDA vs CPSC vs NHTSA)

More recalls

The text below is copied from the official FDA (device) record. If anything differs, the official notice is correct.

Company / brand
Howmedica Osteonics Corp.
Product
TS Femoral Prep Tray, Triathlon. Catalog Number: 6543-6-507.
Reason / hazard
Reprocessing instructions provided in LSTPI-B specify a sterilization dry time of 30 minutes for the trays sterilized in blue wrap. When blue wrap is used, the 30-minute dry time may lead to a small amount of residual moisture remaining after sterilization.
What to do (remedy)

How to claim your remedy

Classification
Class II
Status
Ongoing
Distribution / sold at
These trays are provided by sale or loan. Loans are provided through Distributor. Foreign Consignees will be identified by the countries impacted. 158 trays have been distributed OUS.
Recall initiated
2026-09-16
Code info
Catalog Number: 6543-6-507. GTIN: 07613327557138. All lot numbers used prior to the correction date of May 8, 2026 are impacted.
Product quantity
582 trays
Voluntary mandated
Voluntary: Firm initiated
City
Mahwah
State
NJ
Country
United States
Event id
99883
Free daily digest + "My products" watchlist

Get the daily digest, and tell us what to watch. Everything is optional except your email.

Free allergen alerts:

Categories:

Privacy: we store only what you enter here, use it only to send recall alerts you asked for, never sell or share it, and you can unsubscribe anytime. Double opt-in: we will email you a confirmation link before sending anything; alerts start only after you confirm. Every email has unsubscribe and manage-watchlist links.