Howmedica Osteonics Corp.: TS Misc. Instruments Tray, Triathlon. Catalog Number: 6543-6-519.
- Recall status
- Ongoing (still active)
- Severity
- Class II: could cause temporary or reversible health problems What recall classes mean
- Where sold
- These trays are provided by sale or loan. Loans are provided through Distributor. Foreign Consignees will be identified by the countries impacted. 158 trays have been distributed OUS.
Common questions
- Is Howmedica Osteonics TS Misc. Instruments Tray Triathlon recalled?
- Yes. FDA (device) recalled Howmedica Osteonics TS Misc. Instruments Tray Triathlon on 2026-10-21. Recall status: Ongoing (still active).
- What is the hazard?
- Reprocessing instructions provided in LSTPI-B specify a sterilization dry time of 30 minutes for the trays sterilized in blue wrap. When blue wrap is used, the 30-minute dry time may lead to a small amount of residual moisture remaining after sterilization.
- Where was it sold?
- These trays are provided by sale or loan. Loans are provided through Distributor. Foreign Consignees will be identified by the countries impacted. 158 trays have been distributed OUS.
Read the official notice at FDA (device) →
Is Howmedica Osteonics recalled?
More recalls
- Howmedica Osteonics Corp.: Triathlon Hinge Femoral Prep Tray. Catalog Number
- Howmedica Osteonics Corp.: TS Femoral Prep Tray, Triathlon. Catalog Number
- Howmedica Osteonics Corp.: TS Universal Baseplate Finishing Tray, Triathlon.
- Howmedica Osteonics Corp.: SCORPIO C-DOME PATELLA 73-2710 73-2910 SCORPIO U-DOME
- Howmedica Osteonics Corp.: SCORPIO + MOBILE BEARING PCS 78-5-0910 SYSTEM ONE
The text below is copied from the official FDA (device) record. If anything differs, the official notice is correct.
- Company / brand
- Howmedica Osteonics Corp.
- Product
- TS Misc. Instruments Tray, Triathlon. Catalog Number: 6543-6-519.
- Reason / hazard
- Reprocessing instructions provided in LSTPI-B specify a sterilization dry time of 30 minutes for the trays sterilized in blue wrap. When blue wrap is used, the 30-minute dry time may lead to a small amount of residual moisture remaining after sterilization.
- What to do (remedy)
- Classification
- Class II
- Status
- Ongoing
- Distribution / sold at
- These trays are provided by sale or loan. Loans are provided through Distributor. Foreign Consignees will be identified by the countries impacted. 158 trays have been distributed OUS.
- Recall initiated
- 2026-09-16
- Code info
- Catalog Number: 6543-6-519. GTIN: 07613327555127. All lot numbers used prior to the correction date of May 8, 2026 are impacted.
- Product quantity
- 577 trays
- Voluntary mandated
- Voluntary: Firm initiated
- City
- Mahwah
- State
- NJ
- Country
- United States
- Event id
- 99883