Howmedica Osteonics Corp.: MICS3 Angled Sagittal Saw Attachment; Part Number: 210490
Read the official notice at FDA (device) →
Get alerts for Howmedica Osteonics
The text below is copied from the official FDA (device) record. If anything differs, the official notice is correct.
- Company / brand
- Howmedica Osteonics Corp.
- Product
- MICS3 Angled Sagittal Saw Attachment; Part Number: 210490
- Reason / hazard
- A potential issue was identified with the torque strength on the external screws of the MICS3 Angled Sagittal Saw Attachment. These screws, which secure the cover to the Angled Saw Housing, may loosen during use and detach from the attachment.
- Classification
- Class II
- Status
- Ongoing
- Distribution / sold at
- Domestic: Nationwide Distribution; International: Australia, Germany, Belgium, France, Spain UK, New Zealand;
- Recall initiated
- 2026-02-02
- Code info
- Part Number: 210490; UDI: (01)07613327523874(10)TU6563205RP; Lot numbers: ALL;
- Product quantity
- 1098 units
- Voluntary mandated
- Voluntary: Firm initiated
- City
- Mahwah
- State
- NJ
- Country
- United States
- Event id
- 98411