PHILIPS MEDICAL SYSTEMS NEDERLAND B.V.: Philips Azurion 5M12, Model Numbers: 722227 with Software release R2.2.0, R2.2.1, R2.2.3, R2.2.5 and R2.2.6 with the Roadmap Pro Software k
Read the official notice at FDA (device) →
Get alerts for Philips Medical Systems Nederland B.v
The text below is copied from the official FDA (device) record. If anything differs, the official notice is correct.
- Company / brand
- PHILIPS MEDICAL SYSTEMS NEDERLAND B.V.
- Product
- Philips Azurion 5M12, Model Numbers: 722227 with Software release R2.2.0, R2.2.1, R2.2.3, R2.2.5 and R2.2.6 with the Roadmap Pro Software key
- Reason / hazard
- During the device phase of Roadmap Pro, a user may experience a dark image that is not clinically usable. The dark image may result after the user performs a pedal tap (shortly touching the footswitch pedal without generating an Xray image). The pedal tap may cause a dark image with the next fluoroscopy run.
- Classification
- Class II
- Status
- Ongoing
- Distribution / sold at
- US Nationwide distribution.
- Recall initiated
- 2023-12-07
- Code info
- UDI (01)00884838099227(21)88, (01)00884838099227(21)135, (01)00884838099227(21)63.
- Product quantity
- 3 units
- City
- Best
- Country
- Netherlands
- Event id
- 98977