PHILIPS MEDICAL SYSTEMS NEDERLAND B.V.: Philips Azurion S. Model Number: 722285.
- Recall status
- Ongoing (still active)
- Severity
- Class II: could cause temporary or reversible health problems What recall classes mean
- Where sold
- Worldwide distribution - US Nationwide and the countries of Afghanistan, Albania, Algeria, Argentina, Armenia, Australia, Austria, Azerbaijan, Bahrain, Bangladesh, Belarus, Belgium, Bolivia, Bosnia...
Common questions
- Is Philips Medical Systems Nederland B.v Azurion S recalled?
- Yes. FDA (device) recalled Philips Medical Systems Nederland B.v Azurion S on 2026-11-04. Recall status: Ongoing (still active).
- What is the hazard?
- Solid-State Drives (SSDs) that are not specified in the Medical Device File may have been installed in various PCs. These SSDs may have different performance characteristics, which may include reduced reliability over time such as loss of imaging functionality or motorized movement, a blank or black display, and/or...
- Where was it sold?
- Worldwide distribution - US Nationwide and the countries of Afghanistan, Albania, Algeria, Argentina, Armenia, Australia, Austria, Azerbaijan, Bahrain, Bangladesh, Belarus, Belgium, Bolivia, Bosnia...
Read the official notice at FDA (device) →
Is Philips Medical Systems Nederland B.v recalled?
More recalls
- PHILIPS MEDICAL SYSTEMS NEDERLAND B.V.: Philips Azurion 7 M20. Model Numbers
- PHILIPS MEDICAL SYSTEMS NEDERLAND B.V.: Philips Azurion 7 M12. Model Numbers
- PHILIPS MEDICAL SYSTEMS NEDERLAND B.V.: Philips Azurion 7 B20. Model Numbers
- PHILIPS MEDICAL SYSTEMS NEDERLAND B.V.: Philips Azurion 7 B12. Model Numbers
- PHILIPS MEDICAL SYSTEMS NEDERLAND B.V.: Philips Azurion 5 M20. Model Numbers
The text below is copied from the official FDA (device) record. If anything differs, the official notice is correct.
- Company / brand
- PHILIPS MEDICAL SYSTEMS NEDERLAND B.V.
- Product
- Philips Azurion S. Model Number: 722285.
- Reason / hazard
- Solid-State Drives (SSDs) that are not specified in the Medical Device File may have been installed in various PCs. These SSDs may have different performance characteristics, which may include reduced reliability over time such as loss of imaging functionality or motorized movement, a blank or black display, and/or the inability to store newly acquired images or access previously acquired images.
- What to do (remedy)
- Classification
- Class II
- Status
- Ongoing
- Distribution / sold at
- Worldwide distribution - US Nationwide and the countries of Afghanistan, Albania, Algeria, Argentina, Armenia, Australia, Austria, Azerbaijan, Bahrain, Bangladesh, Belarus, Belgium, Bolivia, Bosnia and Herzegovina, Botswana, Brazil, Brunei Darussal, Bulgaria, Cambodia, Canada, Cayman Islands, Chile, China, Colombia, Costa Rica, C¿te D'Ivoire, Croatia, Cyprus, Czech Republic, Denmark, Dominican Republic, Ecuador, Egypt, Estonia, Ethiopia, Finland, France, Germany, Greece, Hong Kong, Hungary, Iceland, India, Indonesia, Iran, Iraq, Ireland, Israel, Italy, Japan, Jordan, Kazakhstan, Kenya, Korea, South, Kosovo, Kuwait, Kyrgyzstan, Latvia, Lebanon, Libya, Lithuania, Macao, Macedonia, Malaysia, Maldives, Malta, Mauritania, Mauritius, Mexico, Moldova, Mongolia, Morocco, Myanmar, Nepal, Netherlands, New Zealand, Nicaragua, Nigeria, Norway, Oman, Pakistan, Palestine, Panama, Peru, Philippines, Poland, Portugal, Qatar, Romania, Russian Federation, Saudi Arabia, Senegal, Serbia, Singapore, Slovakia, Slovenia, South Africa, Spain, Sri Lanka, Sweden, Switzerland, Syrian Arab Republic, Taiwan, Tanzania, Thailand, Tunisia, Turkey, Turkmenistan, Ukraine, United Arab Emirates, United Kingdom, Uruguay, Uzbekistan, Vietnam, Yemen.
- Recall initiated
- 2026-10-06
- Code info
- 1. Model Number: 722285. UDI Number: (01)00884838125681(21)7. Serial Number: 7.
- Product quantity
- 1 unit
- Voluntary mandated
- Voluntary: Firm initiated
- City
- Best
- Country
- Netherlands
- Event id
- 99939