I.T.S. GmbH: Pilonplate with Angular Stability, with the below descriptions and corresponding article numbers. 1. Anterolateral Tibia Plate; Article
Read the official notice at FDA (device) →
The text below is copied from the official FDA (device) record. If anything differs, the official notice is correct.
- Company / brand
- I.T.S. GmbH
- Product
- Pilonplate with Angular Stability, with the below descriptions and corresponding article numbers. 1. Anterolateral Tibia Plate; Article Numbers: 21311-11, 21311-15, 21311-7, 21312-11, 21312-15, 21312-7. 2. Pilon Plate 4mm; Article Numbers: 21094-10, 21094-4, 21094-6, 21094-8. 3. Pilon Plate, Small; Article Numbers: 21093-4, 21093-6.
- Reason / hazard
- Updated MRI safety testing demonstrates higher RF-induced temperature increases under certain MRI conditions than previously reflected in the Instructions for Use (IFU).
- Classification
- Class II
- Status
- Ongoing
- Distribution / sold at
- US Nationwide distribution in the states of AL, AZ, BVI, CA, CO, FL, GA, IL, IN, KS, KY, LA, MA, MD, MI, MN, MO, MS, MT, NE, NJ, NM, NV, OH, OK, PA, SC, TX, USVI, WV, W.
- Recall initiated
- 2026-01-23
- Code info
- 1. Article Number (Lot Numbers): 21311-11 (26/F62573), 21311-15 (26/F65048, 26/F67521), 21311-7 (26/F66165, 26/F67825), 21312-11 (26/F65049, 26/F67826), 21312-15 (26/F66166), 21312-7 (26/F65643, 26/F67827). 2. Article Number (Lot Numbers): 21094-10 (643/032304, 643/082309), 21094-4 (643/0572, 643/0571), 21094-6 (643/112202), 21094-8 (643/012309). 3.Article Number (Lot Numbers): 21093-4 (679/032311, 679/112331), 21093-6 (679/112332).
- Product quantity
- 101 units
- Voluntary mandated