I.T.S. GmbH: I.T.S. DUL (Distal Ulna Locking) System and UOL (Ulna Osteotomy Locking) System with the below descriptions and corresponding article number
Read the official notice at FDA (device) →
The text below is copied from the official FDA (device) record. If anything differs, the official notice is correct.
- Company / brand
- I.T.S. GmbH
- Product
- I.T.S. DUL (Distal Ulna Locking) System and UOL (Ulna Osteotomy Locking) System with the below descriptions and corresponding article numbers. 1. I.T.S. DUL (Distal Ulna Locking) System, Distal Ulna Plate; Article Numbers: 21232-3, 21232-4, 21232-6, 21233-3, 21233-4, 21233-6, 21234-4, 21235-4. 2. I.T.S. UOL (Ulna Osteotomy Locking) System, Ulna Osteotomy Plate; Article Numbers: 21231-6. 3. UOL (Ulna Osteotomy Locking) System, Cancellous Screw, Locking; Article Numbers: 37302-18, 37302-20.
- Reason / hazard
- Updated MRI safety testing demonstrates higher RF-induced temperature increases under certain MRI conditions than previously reflected in the Instructions for Use (IFU).
- Classification
- Class II
- Status
- Ongoing
- Distribution / sold at
- US Nationwide distribution in the states of AL, AZ, BVI, CA, CO, FL, GA, IL, IN, KS, KY, LA, MA, MD, MI, MN, MO, MS, MT, NE, NJ, NM, NV, OH, OK, PA, SC, TX, USVI, WV, W.
- Recall initiated
- 2026-01-23
- Code info
- 1. Article Number (Lot Numbers): 21232-3 (26/F65636), 21232-4 (26/F61722, 26/F65637), 21232-6 (26/F62564, 26/F65638), 21233-3 (26/F65639), 21233-4 (26/F65640), 21233-6 (26/F61723, 26/F65046), 21234-4 (26/F65641, 26/F66504), 21235-4 (26/F65642, 26/F67810). 2. Article Number (Lot Numbers): 21231-6 (643/012301, 643/022310, 643/032308, 643/0630, 643/0657, 643/102402, 643/122301). 3. Article Number (Lot Numbers): 37302-18 (18/636447), 37302-20 (18/802189).
- Product quantity
- 120 units
- Voluntary mandated
- Voluntary: Firm initiated
- City
- Lasnitzhohe
- Country
- Austria
- Event id
- 98341