Philips North America Llc: IQon Spectral CT; Product Code (REF): 728332;

Medical DevicesSource: FDA (device) · Date: 2026-02-18

Read the official notice at FDA (device) →

Share on XFacebookRedditEmail

Get alerts for Philips North America

Guides: Which agency handles which recall (FDA vs USDA vs CPSC vs NHTSA)

The text below is copied from the official FDA (device) record. If anything differs, the official notice is correct.

Company / brand
Philips North America Llc
Product
IQon Spectral CT; Product Code (REF): 728332;
Reason / hazard
As part of the CT systems sub assembly, thirty-two (32) fasteners are used to attach the rotor to the bearing within the gantry. One or more of these fasteners may not be torqued to specification. If multiple fasteners are not torqued to specification, the rotor or rotor parts may become unsecured or displaced. While there has been no reported or observed event of expelled parts during gantry rotation, unsecured rotor parts may potentially be expelled.
Classification
Class II
Status
Ongoing
Distribution / sold at
Worldwide - US Nationwide distribution in the states of AL, AR, CA, CO, CT, FL, GA, HI, IL, IN, KY, LA, MA, MD, MN, NC, NY, OH, OR, PA, TX, VT, WA, WV and the countries of Australia, Austria, Bahrain, Belgium, Brazil, Bulgaria, Chile, China, Costa Rica, Croatia, Czech Republic, Denmark, Ecuador, Egypt, El Salvador, France, Germany, Hong Kong, India, Indonesia, Iran, Iraq, Ireland, Israel, Italy, Japan, Jordan, Kazakhstan, Kuwait, Lithuania, Macao, Mexico, Netherlands, New Zealand, Norway, Panama, Philippines, Poland, Portugal, R¿union, Russian Fed., Saudi Arabia, Serbia, Singapore, South Africa, South Korea, Spain, Switzerland, Taiwan, Thailand, Turkmenistan, United Kingdom, Uzbekistan, Vietnam.
Recall initiated
2026-01-07
Code info
Product Code (REF): 728332; UDI-DI: 00884838059542; Serial Numbers: 60128, 60141, 60151, 60161, 60171, 60181, 60129, 60142, 60152, 60162, 60172, 60182, 60130, 60143, 60153, 60163, 60173, 60132, 60144, 60154, 60164, 60174, 60133, 60145, 60155, 60165, 60175, 60134, 60146, 60156, 60166, 60176, 60136, 60147, 60157, 60167, 60177, 60137, 60148, 60158, 60168, 60178, 60138, 60149, 60159, 60169, 60179, 60139, 60150, 60160, 60170, 60180,
Product quantity
52 units
Voluntary mandated
Voluntary: Firm initiated
City
Cambridge
State
MA
Country
United States
Event id
98248
Free daily digest + "My products" watchlist

Get the daily digest, and tell us what to watch. Everything is optional except your email.

Free allergen alerts:

Categories:

Privacy: we store only what you enter here, use it only to send recall alerts you asked for, never sell or share it, and you can unsubscribe anytime. Double opt-in: we will email you a confirmation link before sending anything; alerts start only after you confirm. Every email has unsubscribe and manage-watchlist links.