Philips North America: Upgrade to MR 7700; Model number (REF): 782130; Software Versions: R11, R12 and R13;

Medical DevicesSource: FDA (device) · Date: 2026-10-07

Read the official notice at FDA (device) →

Share on XFacebookRedditEmail

Get alerts for Philips North America

Guides: Which agency handles which recall (FDA vs USDA vs CPSC vs NHTSA)

The text below is copied from the official FDA (device) record. If anything differs, the official notice is correct.

Company / brand
Philips North America
Product
Upgrade to MR 7700; Model number (REF): 782130; Software Versions: R11, R12 and R13;
Reason / hazard
When exporting diffusion tensor imaging (DTI) scans from a MR system using the Enhanced MR DICOM format, the exports contain incorrect data (diffusion directions). If the Enhanced MR DICOM export is used in conjunction with third-party software for tractography (e.g. for mapping nerve fiber pathways in the brain), the software will not be able to correctly process the DTI data for reconstruction.
Classification
Class II
Status
Ongoing
Distribution / sold at
Worldwide distribution - Nationwide and the countries of Algeria, Argentina, Australia, Austria, Bangladesh, Belgium, Brazil, Canada, Chile, China, Colombia, Cyprus, Czech Republic, Denmark, Dominican Rep, Ecuador, Egypt, Equator. Guinea, Finland, France, French Guiana, Georgia, Germany, Ghana, Greece, Honduras, Hong Kong, Hungary, India, Indonesia, Iran, Iraq, Ireland, Israel, Italy, Japan, Jordan, Kazakhstan, Kenya, Kyrgyzstan, Latvia, Lebanon, Luxembourg, Malaysia, Malta, Martinique, Mauritius, Mexico, Moldova, Monaco, Morocco, Myanmar, Namibia, Nepal, Netherlands, New Zealand, Nicaragua, Norway, Pakistan, Palestine, Peru, Philippines, Poland, Portugal, Qatar, Romania, Russian Fed., Saudi Arabia, Serbia, Singapore, Slovakia, South Africa, South Korea, Spain, Sweden, Switzerland, Syria, Taiwan, Thailand, Tunisia, T¿rkiye, United Kingdom, Uruguay, Utd.Arab.Emir., Uzbekistan, Vietnam.
Recall initiated
2026-08-28
Code info
Model number (REF): 782130; UDI-DI: 00884838104402; Serial Numbers: 42002, 42030, 42037, 42066, 42104, 42151, 42163, 42164, 42211, 42219, 42248, 42376, 45001, 45007, 45043, 45044, 45052, 45179, 45236, 45263, 45552, 71060, 71262, 71279, 71406, 71532, 71565, 71580, 71616, 71907;
Product quantity
30
Voluntary mandated
Voluntary: Firm initiated
City
Cambridge
State
MA
Country
United States
Event id
99834
Free daily digest + "My products" watchlist

Get the daily digest, and tell us what to watch. Everything is optional except your email.

Free allergen alerts:

Categories:

Privacy: we store only what you enter here, use it only to send recall alerts you asked for, never sell or share it, and you can unsubscribe anytime. Double opt-in: we will email you a confirmation link before sending anything; alerts start only after you confirm. Every email has unsubscribe and manage-watchlist links.