BioMerieux SA: EMAG system Model/Catalog Number: 418591/418591U Product Description: Nucleic acid sample preparation instrument IVD, automated
- Recall status
- Ongoing (still active)
- Severity
- Class II: could cause temporary or reversible health problems What recall classes mean
- Where sold
- Worldwide - US Nationwide distribution in the states of CA, IN, KS, LA, MN, MO, NY, OH, and TX. The countries of Canada, United Kingdom, Sweden, France, Norway, Netherlands, Belgium, Denmark, Germany...
Common questions
- Is BioMerieux SA EMAG system recalled?
- Yes. FDA (device) recalled BioMerieux SA EMAG system on 2026-11-04. Recall status: Ongoing (still active).
- What is the hazard?
- Delayed or no result when analyzing clinical samples with EMAG system (ref: 418591/418591U) due to a randomly occurring software error, Error 10003 Heater timeout after updating to recent firmware SP4 Patch 2
- Where was it sold?
- Worldwide - US Nationwide distribution in the states of CA, IN, KS, LA, MN, MO, NY, OH, and TX. The countries of Canada, United Kingdom, Sweden, France, Norway, Netherlands, Belgium, Denmark, Germany...
Read the official notice at FDA (device) →
More recalls
- Medline: Beds, Rails, Mattresses, and accessories
- ARJOHUNTLEIGH POLSKA Sp. z.o.o.: Carevo, Height-adjustable shower trolley, Model
- Stryker Leibinger GmbH & Co. KG: Precision Pointer Electromagnetic
- Boston Scientific Neuromodulation Corporation: Nalu TD2 Therapy Disc with
- Pentax of America Inc: PENTAX Medical Water bottle assembly. Model Number.
The text below is copied from the official FDA (device) record. If anything differs, the official notice is correct.
- Company / brand
- BioMerieux SA
- Product
- EMAG system Model/Catalog Number: 418591/418591U Product Description: Nucleic acid sample preparation instrument IVD, automated
- Reason / hazard
- Delayed or no result when analyzing clinical samples with EMAG system (ref: 418591/418591U) due to a randomly occurring software error, Error 10003 Heater timeout after updating to recent firmware SP4 Patch 2
- What to do (remedy)
- Classification
- Class II
- Status
- Ongoing
- Distribution / sold at
- Worldwide - US Nationwide distribution in the states of CA, IN, KS, LA, MN, MO, NY, OH, and TX. The countries of Canada, United Kingdom, Sweden, France, Norway, Netherlands, Belgium, Denmark, Germany, Austria.
- Recall initiated
- 2026-09-11
- Code info
- UDI-DI 357302BUDI000544TP Serial Numbers IM03343 IM03283 IM03104 IM03329 IM03376 IM03474 IM03321 IM03071 IM03175 IM03194 IM03515 IM03516 IM03240 IM03478 IM03479 IM03482 IM03034 IM03092 IM03378 IM03414 IM03177 IM03193 IM03506 IM3054 IM03439 IM03350 IM03380 IM03131 IM03148 IM03153 IM03275 IM03314 IM03351
- Product quantity
- 33
- Voluntary mandated
- Voluntary: Firm initiated
- City
- Marcy L'Etoile
- Country
- France
- Event id
- 99931