⚠ RecallRadar📷 ScanGuidesRecentOfficial U.S. recall notices, restated plainly

Boston Scientific Neuromodulation Corporation: Nalu TD2 Therapy Disc with firmware, REF: 34014-001, and TD2 Patient Kit, REF: 74011, part of the Nalu Neurostimulation System

Medical DevicesSource: FDA (device) · Date: 2026-11-04

Recall status
Ongoing (still active)
Severity
Class II: could cause temporary or reversible health problems What recall classes mean
Where sold
US Nationwide distribution in the states of PA, FL, OR, AZ, OH, SC, TX, OR, OK, MI, CA, NJ.

Common questions

Is Boston Scientific Neuromodulation Nalu TD2 Therapy Disc with firmware REF recalled?
Yes. FDA (device) recalled Boston Scientific Neuromodulation Nalu TD2 Therapy Disc with firmware REF on 2026-11-04. Recall status: Ongoing (still active).
What is the hazard?
Therapy Disc (TD) that powers and sends stimulation signals to an Implantable Pulse Generator (IPG), part of a neurostimulation system, has a firmware issue that causes the TD to attempt to restart, then fail, leading it to become non-functional, and if a secondary TD is unavailable, this may lead IPG to become...
Where was it sold?
US Nationwide distribution in the states of PA, FL, OR, AZ, OH, SC, TX, OR, OK, MI, CA, NJ.

Read the official notice at FDA (device) →

Share on XFacebookRedditEmail

Is Boston Scientific Neuromodulation recalled?

Guides: Which agency handles which recall (FDA vs USDA vs CPSC vs NHTSA)

More recalls

The text below is copied from the official FDA (device) record. If anything differs, the official notice is correct.

Company / brand
Boston Scientific Neuromodulation Corporation
Product
Nalu TD2 Therapy Disc with firmware, REF: 34014-001, and TD2 Patient Kit, REF: 74011, part of the Nalu Neurostimulation System
Reason / hazard
Therapy Disc (TD) that powers and sends stimulation signals to an Implantable Pulse Generator (IPG), part of a neurostimulation system, has a firmware issue that causes the TD to attempt to restart, then fail, leading it to become non-functional, and if a secondary TD is unavailable, this may lead IPG to become non-functional, thus unable to inhibit transmission of pain signals.
What to do (remedy)

How to claim your remedy

Classification
Class II
Status
Ongoing
Distribution / sold at
US Nationwide distribution in the states of PA, FL, OR, AZ, OH, SC, TX, OR, OK, MI, CA, NJ.
Recall initiated
2026-09-30
Code info
Firmware versions v2.0.71 and v2.0.77 REF/UDI-DI/Serial Numbers: 34014-001/00812537037134/ 100855, 100877, 100883, 100884, 100886, 100888, 100898, 100903, 100908, 100911, 100916, 100932, 101103, 100902; 74011/00812537037127, 00812537036977/101439, 100913, 100949, 101159, 101162, 101163, 101164, 101173, 101175, 101177, 101186, 101187, 101188, 101189, 101192, 101200, 101205, 101208, 101213, 101214, 101218, 101219, 101221, 101222, 101240, 101245, 101247, 101249, 101256, 101257, 101259, 101260, 101261, 101264, 101267, 101268, 101270, 101272, 101273, 101277, 101278, 101280, 101283, 101284, 101290, 101297, 101298, 101301, 101305, 101307, 101310, 101311, 101312, 101314, 101315, 101326, 101330, 101331, 101334, 101338, 101344, 101345, 101346, 101356, 101357, 101358, 101359, 101363, 101364, 101367, 101375, 101376, 101377, 101378, 101381, 101383, 101387, 101388, 101392, 101400, 101405, 101408, 101410, 101414, 101415, 101417, 101418, 101419, 101429, 101431, 101432, 101436, 101440, 101441, 101443, 101447, 101448, 101450, 101451, 101466, 101490, 101493, 101504, 101510, 101511, 101514, 101520, 101527, 101528, 101534, 101535, 101553, 101557, 101567, 101571, 101576, 101599, 101617, 101623, 101656
Product quantity
134
Voluntary mandated
Voluntary: Firm initiated
City
Valencia
State
CA
Country
United States
Event id
99785
Free daily digest + "My products" watchlist

Get the daily digest, and tell us what to watch. Everything is optional except your email.

Free allergen alerts:

Categories:

Privacy: we store only what you enter here, use it only to send recall alerts you asked for, never sell or share it, and you can unsubscribe anytime. Double opt-in: we will email you a confirmation link before sending anything; alerts start only after you confirm. Every email has unsubscribe and manage-watchlist links.