TECHNO-PATH MANUFACTURING LTD.: Techno-Path Multichem AE Unassayed Level 3, REF:AE623X

Medical DevicesSource: FDA (device) · Date: 2026-10-28

Recall status
Ongoing (still active)
Severity
Class II: could cause temporary or reversible health problems What recall classes mean
Where sold
US Nationwide distribution in the state of Massachusetts.

Common questions

Is Techno-path Manufacturing Multichem AE Unassayed Level 3 REF:AE623X recalled?
Yes. FDA (device) recalled Techno-path Manufacturing Multichem AE Unassayed Level 3 REF:AE623X on 2026-10-28. Recall status: Ongoing (still active).
What is the hazard?
A routine stability studies have identified a stability drift affecting the Ammonia (NH4+) analyte in specific lots of Multichem AE quality control material. Additional testing confirmed that the labelled shelf-life performance claim for the Ammonia analyte can no longer be supported for the affected lots. This...
Where was it sold?
US Nationwide distribution in the state of Massachusetts.

Read the official notice at FDA (device) →

Share on XFacebookRedditEmail

Is Techno-path Manufacturing recalled?

Guides: Which agency handles which recall (FDA vs USDA vs CPSC vs NHTSA)

More recalls

The text below is copied from the official FDA (device) record. If anything differs, the official notice is correct.

Company / brand
TECHNO-PATH MANUFACTURING LTD.
Product
Techno-Path Multichem AE Unassayed Level 3, REF:AE623X
Reason / hazard
A routine stability studies have identified a stability drift affecting the Ammonia (NH4+) analyte in specific lots of Multichem AE quality control material. Additional testing confirmed that the labelled shelf-life performance claim for the Ammonia analyte can no longer be supported for the affected lots. This issue is limited to the Ammonia analyte only. The impacted device may cause delay of issuing results to the physician. Delayed recognition and diagnosis of elevated levels may impact in timely treatment of life-threatening levels described within medical toxicity intervention decision points.
What to do (remedy)

How to claim your remedy

Classification
Class II
Status
Ongoing
Distribution / sold at
US Nationwide distribution in the state of Massachusetts.
Recall initiated
2026-08-14
Code info
UDI:(01)05391523442824/-Lot # 65705263
Product quantity
832 units
Voluntary mandated
Voluntary: Firm initiated
City
Ballina Tipperary
Country
Ireland
Event id
99724
Free daily digest + "My products" watchlist

Get the daily digest, and tell us what to watch. Everything is optional except your email.

Free allergen alerts:

Categories:

Privacy: we store only what you enter here, use it only to send recall alerts you asked for, never sell or share it, and you can unsubscribe anytime. Double opt-in: we will email you a confirmation link before sending anything; alerts start only after you confirm. Every email has unsubscribe and manage-watchlist links.