TECHNO-PATH MANUFACTURING LTD.: Techno-Path Multichem AE Unassayed Level 2 REF: AE622X
- Recall status
- Ongoing (still active)
- Severity
- Class II: could cause temporary or reversible health problems What recall classes mean
- Where sold
- US Nationwide distribution in the state of Massachusetts.
Common questions
- Is Techno-path Manufacturing Multichem AE Unassayed Level 2 REF: AE622X recalled?
- Yes. FDA (device) recalled Techno-path Manufacturing Multichem AE Unassayed Level 2 REF: AE622X on 2026-10-28. Recall status: Ongoing (still active).
- What is the hazard?
- A routine stability studies have identified a stability drift affecting the Ammonia (NH4+) analyte in specific lots of Multichem AE quality control material. Additional testing confirmed that the labelled shelf-life performance claim for the Ammonia analyte can no longer be supported for the affected lots. This...
- Where was it sold?
- US Nationwide distribution in the state of Massachusetts.
Read the official notice at FDA (device) →
Is Techno-path Manufacturing recalled?
More recalls
- TECHNO-PATH MANUFACTURING LTD.: Techno-Path Multichem AE Unassayed Level 3
- TECHNO-PATH MANUFACTURING LTD.: Techno-Path Multichem AE Unassayed Level 1
- TECHNO-PATH MANUFACTURING LTD.: Multichem ID-B. Model Number: SR103B. Intended
- TECHNO-PATH MANUFACTURING LTD.: Multichem ID-B. Model/Catalog Number: SR103B
- Medline: Beds, Rails, Mattresses, and accessories
The text below is copied from the official FDA (device) record. If anything differs, the official notice is correct.
- Company / brand
- TECHNO-PATH MANUFACTURING LTD.
- Product
- Techno-Path Multichem AE Unassayed Level 2 REF: AE622X
- Reason / hazard
- A routine stability studies have identified a stability drift affecting the Ammonia (NH4+) analyte in specific lots of Multichem AE quality control material. Additional testing confirmed that the labelled shelf-life performance claim for the Ammonia analyte can no longer be supported for the affected lots. This issue is limited to the Ammonia analyte only. The impacted device may cause delay of issuing results to the physician. Delayed recognition and diagnosis of elevated levels may impact in timely treatment of life-threatening levels described within medical toxicity intervention decision points.
- What to do (remedy)
- Classification
- Class II
- Status
- Ongoing
- Distribution / sold at
- US Nationwide distribution in the state of Massachusetts.
- Recall initiated
- 2026-08-14
- Code info
- UDI:(01)05391523442817/Lot # 65705262
- Product quantity
- 123 units
- Voluntary mandated
- Voluntary: Firm initiated
- City
- Ballina Tipperary
- Country
- Ireland
- Event id
- 99724