Medline Industries, LP: Medline Basic Homecare Beds and Medlite Homecare Beds and Accessories, Hand Pendants: MDR510094 PENDANT F/ MDR107002E MDR520094 PENDANT
Read the official notice at FDA (device) →
Get alerts for Medline Industries
The text below is copied from the official FDA (device) record. If anything differs, the official notice is correct.
- Company / brand
- Medline Industries, LP
- Product
- Medline Basic Homecare Beds and Medlite Homecare Beds and Accessories, Hand Pendants: MDR510094 PENDANT F/ MDR107002E MDR520094 PENDANT F/ MDR107003E MDR530094 PENDANT, W/ 9V BATTERY BACKUP, F/ 2E BED MDR540094 PENDANT, W/ 9V BATTERY BACKUP, F/ 3E BED
- Reason / hazard
- Potential fire hazard involving the hand control pendant and associated wiring used with affected beds. Pendant or pendant wiring may overheat, potentially leading to smoking, sparking, melting, burning, or fire. Additional products added to 11/26/2025 recall.
- Classification
- Class I
- Status
- Ongoing
- Distribution / sold at
- Distribution US Nationwide including territories of Puerto Rico and Guam and the countries of Bermuda, Canada, and Panama.
- Recall initiated
- 2026-08-14
- Code info
- MDR510094 UDI-DI 10888277363021 MDR520094 UDI-DI 10884389161692 MDR530094 UDI-DI 10197344119059 MDR540094 UDI-DI 10197344119066 All lots
- Product quantity
- 771,127 total
- Voluntary mandated
- Voluntary: Firm initiated
- City
- Northfield
- State
- IL
- Country
- United States
- Event id
- 99827
Free daily digest + "My products" watchlist