Medtronic MiniMed, Inc.: MiniMed 670G Insulin Pump, Model: MMT-1781, MMT-1782;
Read the official notice at FDA (device) →
Get alerts for Medtronic MiniMed
The text below is copied from the official FDA (device) record. If anything differs, the official notice is correct.
- Company / brand
- Medtronic MiniMed, Inc.
- Product
- MiniMed 670G Insulin Pump, Model: MMT-1781, MMT-1782;
- Reason / hazard
- Loaner insulin pumps may have been cleaned with chemically incompatible products that could damage the pump case or keypad, allow moisture ingress, and result in insulin over- or under-delivery, potentially causing hypoglycemia, hyperglycemia, or diabetic ketoacidosis.
- Classification
- Class II
- Status
- Ongoing
- Distribution / sold at
- International distribution to the countries of Germany, Denmark, Austria, Netherlands, United Kingdom, Italy, Belgium, Ireland, Switzerland, Sweden, Northern Ireland.
- Recall initiated
- 2026-07-13
- Code info
- Model/UDI-DI Serial Numbers: MMT-1781/00763000404536/NG2665223H, NG2748200H, NG2748664H, NG2755237H, NG2716176H, NG3547582H. MMT-1782/00763000404543/NG2768190H
- Product quantity
- 7
- Voluntary mandated
- Voluntary: Firm initiated
- City
- Northridge
- State
- CA
- Country
- United States
- Event id
- 99374