Tobii Dynavox Llc: TD Navio Maxi Speech Generating Device Model/Catalog Number: 1000799

Medical DevicesSource: FDA (device) · Date: 2026-08-26

Read the official notice at FDA (device) →

Share on XFacebookRedditEmail

Guides: Which agency handles which recall (FDA vs USDA vs CPSC vs NHTSA)

The text below is copied from the official FDA (device) record. If anything differs, the official notice is correct.

Company / brand
Tobii Dynavox Llc
Product
TD Navio Maxi Speech Generating Device Model/Catalog Number: 1000799
Reason / hazard
Under certain conditions, the battery mounting tabs may become damaged, allowing the battery to move within the device enclosure, potentially leading to an internal short circuit. This may cause localized heating and gas venting within the device enclosure.
Classification
Class II
Status
Ongoing
Distribution / sold at
US Nationwide distribution.
Recall initiated
2026-07-08
Code info
UDI-DI 073400747602070 Serial Numbers TDNL1-240904000016 through TDNL1-260617070001. Corrected/future production is differentiated by new serialization from TDNL1-260617070001 onward (production batch 20260615, improved design).
Voluntary mandated
Voluntary: Firm initiated
City
Pittsburgh
State
PA
Country
United States
Event id
99410
Free daily digest + "My products" watchlist

Get the daily digest, and tell us what to watch. Everything is optional except your email.

Free allergen alerts:

Categories:

Privacy: we store only what you enter here, use it only to send recall alerts you asked for, never sell or share it, and you can unsubscribe anytime. Double opt-in: we will email you a confirmation link before sending anything; alerts start only after you confirm. Every email has unsubscribe and manage-watchlist links.