United Orthopedic Corp.: Brand Name: U2 Total Knee System Product Name: Femoral valgus offset adaptor, 2mm, 4mm, 6mm, left and Femoral valgus offset adaptor, 2mm, 4

Medical DevicesSource: FDA (device) · Date: 2026-08-12

Read the official notice at FDA (device) →

Share on XFacebookRedditEmail

Get alerts for United Orthopedic

Guides: Which agency handles which recall (FDA vs USDA vs CPSC vs NHTSA)

The text below is copied from the official FDA (device) record. If anything differs, the official notice is correct.

Company / brand
United Orthopedic Corp.
Product
Brand Name: U2 Total Knee System Product Name: Femoral valgus offset adaptor, 2mm, 4mm, 6mm, left and Femoral valgus offset adaptor, 2mm, 4mm, 6mm, right Model/Catalog Number: 9323-5335, 9323-5337, 9323-5339 and 9323-5336, 9323-5338, 9323-5340 Software Version: N/A Product Description: Manual Orthopedic Surgical Instrument Component: No
Reason / hazard
Due to an incomplete process of the component screw by a subcontractor which prevents the properly assembling of stem trail.
Classification
Class II
Status
Ongoing
Distribution / sold at
US Nationwide distribution in the states of Texas and Michigan.
Recall initiated
2026-05-20
Code info
Product Number; UDI-DI Code; Lot Code: Product number 9323-5335; UDI-DI 04719886906554; Lot number 25FB64 Product number 9323-5337; UDI-DI 04719886906578; Lot number 25FB48 Product number 9323-5339; UDI-DI 04719886906592; Lot number 25FB52 Product number 9323-5336; UDI-DI 04719886906561; Lot number 25FB65 Product number 9323-5338; UDI-DI 04719886906585; Lot number 25FB49 Product number 9323-5340; UDI-DI 04719886906608; Lot number 25FB53
Product quantity
2
Voluntary mandated
Voluntary: Firm initiated
City
Kaohsiung
Country
Taiwan
Event id
99026
Free daily digest + "My products" watchlist

Get the daily digest, and tell us what to watch. Everything is optional except your email.

Free allergen alerts:

Categories:

Privacy: we store only what you enter here, use it only to send recall alerts you asked for, never sell or share it, and you can unsubscribe anytime. Double opt-in: we will email you a confirmation link before sending anything; alerts start only after you confirm. Every email has unsubscribe and manage-watchlist links.