GE Medical Systems Information Technologies Inc: GE Healthcare Portrait Core Services SW V 1.2.0, REF 5700592. patient monitoring system
Read the official notice at FDA (device) →
Get alerts for GE Medical Systems Information Technologies
The text below is copied from the official FDA (device) record. If anything differs, the official notice is correct.
- Company / brand
- GE Medical Systems Information Technologies Inc
- Product
- GE Healthcare Portrait Core Services SW V 1.2.0, REF 5700592. patient monitoring system
- Reason / hazard
- A software vulnerability may allow unauthorized persons to gain privileged access to the device, which could result in loss of patient monitoring, missed patient alarms, and/or access to Protected Health Information (PHI).
- Classification
- Class II
- Status
- Ongoing
- Distribution / sold at
- Worldwide - US Nationwide distribution in the state of MI, and the countries of Czech Republic, Finland, Germany, Netherland, Qatar, UAE.
- Recall initiated
- 2026-07-10
- Code info
- GTIN 00198953056752, Lot/Serial Numbers: SWT25510001TA, 3M1D391215, 3M1D411081, 3M1D411083, 3M1D3B100D, SWT26110010TA, SWT26070001TA, SWT26120008TA, 3M1D411080, 3M1D411082, 3M1D3B100C.
- Product quantity
- 11 units
- Voluntary mandated
- Voluntary: Firm initiated
- City
- Waukesha
- State
- WI
- Country
- United States
- Event id
- 99475