Beckman Coulter Biomedical GmbH: REMISOL Advance with Software Version 2.5 UDI-DI code: 15099590802073 The Normand Info REMISOL Advance Data Manager collects and manages
Read the official notice at FDA (device) →
The text below is copied from the official FDA (device) record. If anything differs, the official notice is correct.
- Company / brand
- Beckman Coulter Biomedical GmbH
- Product
- REMISOL Advance with Software Version 2.5 UDI-DI code: 15099590802073 The Normand Info REMISOL Advance Data Manager collects and manages data and workflow for 128 connections
- Reason / hazard
- Due to software issues where does not handle combo assay results. The software may replace the final combo test result with one of the sub-results resulting in incorrect reporting of test outcomes.
- Classification
- Class II
- Status
- Ongoing
- Distribution / sold at
- U.S.: FL, GA, IL, MI, and PA O.U.S.: Belgium, France, Hungary, Poland, Taiwan (Province of China), Tunisia, United Arab Emirates, and United Kingdom of Great Britain and Northern Ireland.
- Recall initiated
- 2026-04-29
- Code info
- Software Version 2.5 UDI-DI code: 15099590802073
- Product quantity
- 20 systems
- Voluntary mandated
- Voluntary: Firm initiated
- City
- Munchen
- Country
- Germany
- Event id
- 98927