KENT IMAGING, INC.: SnapshotGLO Wound Imaging Device, Model KB100, KB100AA.
Read the official notice at FDA (device) →
The text below is copied from the official FDA (device) record. If anything differs, the official notice is correct.
- Company / brand
- KENT IMAGING, INC.
- Product
- SnapshotGLO Wound Imaging Device, Model KB100, KB100AA.
- Reason / hazard
- The software controlling the wound imaging device's battery charging circuit was not designed to recognize and recover a battery that has fully discharged into a low-power "shipping mode." When this occurs, an affected device will not resume charging when reconnected to power and becomes temporarily unusable/ unavailable for wound imaging until corrected.
- Classification
- Class II
- Status
- Ongoing
- Distribution / sold at
- US Nationwide distribution in the states of PA, IA, FL, TX, GA, NV, CA, UT, IN, DC, KS, IL, TN, AR, AZ.
- Recall initiated
- 2026-06-30
- Code info
- UDI-DI: B376KB1000, Software Revisions v4.2.0 and v4.1.3
- Product quantity
- 77
- Voluntary mandated
- Voluntary: Firm initiated
- City
- Calgary
- State
- N/A
- Country
- Canada
- Event id
- 99406