Tosoh Bioscience Inc: ST AIA-Pack Testosterone, IN VITRO DIAGNOSTIC USE ONLY for the quantitative measurement of testosterone in human serum on specific Tosoh AIA
Read the official notice at FDA (device) →
The text below is copied from the official FDA (device) record. If anything differs, the official notice is correct.
- Company / brand
- Tosoh Bioscience Inc
- Product
- ST AIA-Pack Testosterone, IN VITRO DIAGNOSTIC USE ONLY for the quantitative measurement of testosterone in human serum on specific Tosoh AIA System analyzers, Model Number 0025204
- Reason / hazard
- The Test cup lot F level 1 low-quality control results values for the device were lower than expected on the Automated Immunoassay Analyzer (AIA) which could cause falsely elevated or falsely low total testosterone results. Such erroneous test results may lead to additional testing/diagnostic interventions and in limited cases, may lead to a delay in diagnosis and/or initiation of unnecessary or incorrect clinical treatment measures.
- Classification
- Class II
- Status
- Ongoing
- Distribution / sold at
- Global Distribution. US Nationwide.
- Recall initiated
- 2026-06-12
- Code info
- UDI-DI: 14560189210032; Lot Numbers: FX12404, FY12407, FY12406, FZ12408
- Product quantity
- 310 units
- Voluntary mandated
- Voluntary: Firm initiated
- City
- Grove City
- State
- OH
- Country
- United States
- Event id
- 99210