Daavlin Distributing Company: Photera 600, phototherapy device; Ultraviolet Lamp For Dermatologic Disorders
Read the official notice at FDA (device) →
Get alerts for Daavlin Distributing
The text below is copied from the official FDA (device) record. If anything differs, the official notice is correct.
- Company / brand
- Daavlin Distributing Company
- Product
- Photera 600, phototherapy device; Ultraviolet Lamp For Dermatologic Disorders
- Reason / hazard
- Calibrated Output was not calculating correctly - UV Sensor GAIN being applied was incorrect - GAIN seemed to be "4/5" instead of "7" (counts were at ~1000 instead of ~5000).
- Classification
- Class II
- Status
- Ongoing
- Distribution / sold at
- US Nationwide distribution.
- Recall initiated
- 2026-03-05
- Code info
- Serial Numbers: 6-09033NB4CL, 6-09037NB4CL, 6-09031NB4CL, 6-09538NB4CL, 6-09038NB4CL, 6-09039NB4CL, 6-09040NB4CL, 6-09041NB4CL, 6-09035NB4CL, 6-09034NB4CL, 6-09030NB4CL, 6-09026NB4CL, 6-09023NB4CL, 6-09028NB4CL, 6-09025NB4CL, 6-09043NB4CL, 6-09029NB4CL, 6-09044NB4CL, 6-09022NB4CL, 6-09027NB4CL, 6-09036NB4CL, 6-09045NB4CL, 6-09024NB4CL, 6-07827NB4CL, 6-09539NB4CL, 6-09042NB4CL.
- Product quantity
- 26 units
- Voluntary mandated
- Voluntary: Firm initiated
- City
- Bryan
- State
- OH
- Country
- United States
- Event id
- 99331