Maquet Cardiovascular, LLC: The Vasoview Hemopro 2 Endoscopic Vessel Harvesting System (Model VH-4000; UDI 00607567700406) is a sterile, single-use Class II device tha
Read the official notice at FDA (device) →
The text below is copied from the official FDA (device) record. If anything differs, the official notice is correct.
- Company / brand
- Maquet Cardiovascular, LLC
- Product
- The Vasoview Hemopro 2 Endoscopic Vessel Harvesting System (Model VH-4000; UDI 00607567700406) is a sterile, single-use Class II device that was cleared for marketing by FDA under 510(k) number K101274.
- Reason / hazard
- Getinge identified that Vasoview Hemopro 2 devices from Lot 3000547797 were exposed to an out-of-specification sterilization dose during a gamma sterilization run (Run ID 1183-28782A) performed by STERIS at its Whippany, NJ facility between March 31 and April 1 of 2026.
- Classification
- Class II
- Status
- Ongoing
- Distribution / sold at
- US Nationwide distribution in the states of NY, CA, SC, FL, TX, GA, OR.
- Recall initiated
- 2026-06-02
- Code info
- UDI 00607567700406, Serial/Lot Number 3000547797
- Product quantity
- 205 units
- Voluntary mandated
- Voluntary: Firm initiated
- City
- Wayne
- State
- NJ
- Country
- United States
- Event id
- 99259