GE Medical Systems, LLC: Optima XR220amx, Mobile Digital-Ready Radiographic System
Read the official notice at FDA (device) →
Get alerts for GE Medical Systems
The text below is copied from the official FDA (device) record. If anything differs, the official notice is correct.
- Company / brand
- GE Medical Systems, LLC
- Product
- Optima XR220amx, Mobile Digital-Ready Radiographic System
- Reason / hazard
- GE HealthCare has become aware that in certain Optima XR200/220/240amx systems that have had a column replacement completed, functional testing of the safety pawl may not have been performed to confirm proper assembly. If a failure of the internal counterpoise cable occurs, the Column Safety mechanism will not work, causing the arm with X-ray tube and collimator to drop downward on the column track. This could result in patient injury during an exam.
- Classification
- Class II
- Status
- Ongoing
- Distribution / sold at
- Worldwide - US Nationwide and the countries of Brazil, Canada, Korea, Mexico, Saudi Arabia, Spain, UK.
- Recall initiated
- 2026-06-05
- Code info
- UDI-DI: 00840682115261; System IDs: 05249MOB02.OLD, 205820DP1, 228818XR220, 229276M1, 240566P1, 301295OPTIMA2, 303788SBOPT1, 314251P5, 352LEEDP1, 386FLGDPXR3, 417347DPORT1, 512SAMCSB, 516632OPTIMA2, 540981ODP7, 563264OPT, 601445DP1, 615338HDP01, 661949DP2, 713441OPCDX1, 717657OPT1, 718206OPT240A, 727848SB1, 732321DP1, 757736DP3, 813302OPT1, 828213DXP1, 828286OPT1, 843847220D, 850070941, 865545DP3, 915821OPT1, 941342OP221, 972566XR220E, 985447TM4, SA1208MM05, SB2201700085WK, XR1034798WK7, XR1037005WK4
- Product quantity
- 38 units
- Voluntary mandated
- Voluntary: Firm initiated
- City
- Waukesha
- State
- WI
- Country
- United States
- Event id
- 99279
Free daily digest + "My products" watchlist