BD SWITZERLAND SARL: BD Connecta" BD Luer-Lok" 360 REF 394910 UDI-DI code: 00382903949106 BD Connecta" Stopcocks are used to administer fluids or medication

Medical DevicesSource: FDA (device) · Date: 2026-06-24

Read the official notice at FDA (device) →

Share on XFacebookRedditEmail

Get alerts for Bd Switzerland Sarl

Guides: Which agency handles which recall (FDA vs USDA vs CPSC vs NHTSA)

The text below is copied from the official FDA (device) record. If anything differs, the official notice is correct.

Company / brand
BD SWITZERLAND SARL
Product
BD Connecta" BD Luer-Lok" 360 REF 394910 UDI-DI code: 00382903949106 BD Connecta" Stopcocks are used to administer fluids or medication through one or two different ports via an IV cannula or extension tube. The ports can be used to sample blood or for hemodynamic monitoring. When used with IV lipid nutritional products, the stopcock device can be used for up to 24 hours.
Reason / hazard
Due to stopcock issues when using the plastic luer-lok syringe, the connection at the Port C may over thread or continue to spin rather than tightening securely. This behavior has led some clinical users to perceive the connection as unstable.
Classification
Class II
Status
Ongoing
Distribution / sold at
U.S.: AK, AL, AR, AZ, CA, CO, DC, FL, GA, IA, IL, IN, KS, KY, LA, MA, MD, MI, MN, MO, MS, NC, NE, NJ, NM, NY, OH, OK, OR, PA, SC, SD, TN, TX, UT, VA, WA, WI, and WV. O.U.S.: Japan
Recall initiated
2026-05-21
Code info
Catalog/Model Number: 394910 UDI-DI code: 00382903949106 Lot Numbers: 4277968 4277973 4282451 4282461 4289801 4290582 4292904 4295848 4299949 4303841 4305108 4306700 5129635 5129642 5129649 5129650 5129651 5139250 5139251 5139252 5150181 5153940 5164061 5164062 5178078 5181442 5183810 5183816 5183824 5183828 5218287 5218291 5218293 5218299 5218302 5241476 5247729 5247731 5251098 5275601 5275602
Product quantity
2,465,500 units
Voluntary mandated
Voluntary: Firm initiated
City
Eysins
Country
Switzerland
Event id
98855
Free daily digest + "My products" watchlist

Get the daily digest, and tell us what to watch. Everything is optional except your email.

Free allergen alerts:

Categories:

Privacy: we store only what you enter here, use it only to send recall alerts you asked for, never sell or share it, and you can unsubscribe anytime. Double opt-in: we will email you a confirmation link before sending anything; alerts start only after you confirm. Every email has unsubscribe and manage-watchlist links.