ARROW INTERNATIONAL, LLC: Description/REF: HEMODIALYSIS KIT: 2-LUMEN 12 FR X 20 CM/AK-15122-F, AHDC KIT: 3L 12 FR X 16 CM/ASK-12123-UOC1, AHDC KIT: 3-L 12 FR X 16 CM/
Read the official notice at FDA (device) →
Get alerts for Arrow International
The text below is copied from the official FDA (device) record. If anything differs, the official notice is correct.
- Company / brand
- ARROW INTERNATIONAL, LLC
- Product
- Description/REF: HEMODIALYSIS KIT: 2-LUMEN 12 FR X 20 CM/AK-15122-F, AHDC KIT: 3L 12 FR X 16 CM/ASK-12123-UOC1, AHDC KIT: 3-L 12 FR X 16 CM/ASK-12123-UPM, AHDC KIT: 3L 12 FR X 20 CM/ASK-15123-UOC1, AHDC KIT: 3-L 12 FR X 20 CM/ASK-15123-UPM, AHDC KIT: 3-L 12 FR X 16 CM/CDC-12123-P1A, AHDC KIT: 3-L 12 FR X 16 CM/CDC-12123-XP1A, AHDC KIT: 3-L 12 FR X 16 CM/CDC-12123-XPCN1A, AHDC KIT: 3-L 12 FR X 16 CM/CDC-12123-XPN1A, AHDC KIT: 3-L 12 FR X 20 CM/CDC-15123-P1A, AHDC KIT: 3-L 12 FR X 20 CM/CDC-15123-XPCN1A, AHDC KIT: 3-L 12 FR X 20 CM/CDC-15123-XPN1A, AHDC KIT: 3-L 12 FR X 25 CM/CDC-16123-XPCN1A, HEMODIALYSIS KIT 2L: 12 FR X 20CM AGB/AK-25122-CDC, HEMODIALYSIS KIT: 2-LUMEN 14 FR X 20 CM/AK-25142-F, HEMODIALYSIS KIT: 2L 12 FR X 6 IN/ASK-22122-MM1, HEMODIALYSIS KIT: 2L 12 FR X 6 IN/ASK-22122-MM2, AHDC KIT: 2L 14 FR X 15 CM/ASK-22142-UOC1, AHDC AGB KIT: 2-L 14 FR X 15 CM/ASK-22142-UPM, HEMODIALYSIS KIT: 2L 14 FR X 20 CM/ASK-25142-UNC1, AHDC KIT: 2L 14 FR X 20 CM/ASK-25142-UOC1, AHDC AGB KIT: 2L 14 FR X 20 CM/ASK-25142-UPM, HEMODIALYSIS KIT 2L: 14 FR X 25 CM AGB/ASK-26142-UNC1, HEMODIALYSIS KIT: 2LUMEN 12 FR X 16 CM/AU-22122-F, HEMODIALYSIS KIT: 2L 12 FR X 16 CM AGB/CDC-22122-1A, AHDC AGB U-BEND KIT: 2-L 12 FR X 16 CM/CDC-22122-XCN1A-U, AHDC AGB KIT: 2-L 12 FR X 16 CM/CDC-22122-XN1A, AHDC AGB U-BEND KIT: 2-L 12 FR X 16 CM/CDC-22122-XN1A-U, HEMODIALYSIS KIT 2L: 14 FR X 15 CM AGB/CDC-22142-1A, HEMODIALYSIS KIT: 2L 14 FR X 15 CM AGB/CDC-22142-XC1A, AHDC AGB KIT: 2-L 14 FR X 15 CM/CDC-22142-XCN1A, AHDC AGB CURVED KIT: 2-L 14 FR X 15 CM/CDC-22142-XCN1A-C, AHDC AGB KIT: 2-L 14 FR X 15 CM/CDC-22142-XN1A, AHDC AGB CURVED KIT: 2-L 14 FR X 15 CM/CDC-22142-XN1A-C, AHDC AGB U-BEND KIT: 2-L 12 FR X 13 CM/CDC-23122-XCN1A-U, AHDC AGB U-BEND KIT: 2-L 12 FR X 13 CM/CDC-23122-XN1A-U, HEMODIALYSIS KIT: 2L 12 FR X 13 CM AGB/CDC-23122-XU1A, HEMODIALYSIS KIT: 2L 12 FR X 20 CM AGB/CDC-25122-1A, HEMODIALYSIS KIT: 2L 12 FR X 20 CM AGB/CDC-25122-X1A, AHDC AGB KIT: 2-L 12 FR X 20 CM/CDC-25122-XCN1A, AHDC AGB U-BEND KIT: 2-L 12 FR X 20 CM/CDC-25122-XCN1A-U, AHDC AGB KIT: 2-L 12 FR X 20 CM/CDC-25122-XN1A, HEMODIALYSIS KIT: 2L 12 FR X 20 CM AGB/CDC-25122-XU1A, HEMODIALYSIS KIT 2L: 14 FR X 20 CM AGB/CDC-25142-1A, HEMODIALYSIS KIT: 2L 14 FR X 20 CM AGB/CDC-25142-XC1A, AHDC AGB KIT: 2-L 14 FR X 20 CM/CDC-25142-XCN1A, AHDC AGB CURVED KIT: 2-L 14 FR X 20 CM/CDC-25142-XCN1A-C, AHDC AGB KIT: 2-L 14 FR X 20 CM/CDC-25142-XN1A, AHDC AGB CURVED KIT: 2-L 14 FR X 20 CM/CDC-25142-XN1A-C, HEMODIALYSIS KIT: 2L 12 FR X 25 CM AGB/CDC-26122-1A, AHDC AGB KIT: 2-L 12 FR X 25 CM/CDC-26122-XCN1A, AHDC AGB KIT: 2-L 12 FR X 25 CM/CDC-26122-XN1A, HEMODIALYSIS KIT 2L: 14 FR X 25 CM AGB/CDC-26142-1A, HEMODIALYSIS KIT: 2L 14 FR X 25 CM AGB/CDC-26142-X1A, AHDC AGB KIT: 2-L 14 FR X 25 CM/CDC-26142-XCN1A, AHDC AGB KIT: 2-L 14 FR X 25 CM/CDC-26142-XN1A, HEMODIALYSIS KIT: 2-LUMEN 12 FR X 20 CM/AK-25122-F, AHDC AGB KIT: 2-L 12 FR X 16 CM/CDC-22122-XCN1A, LBCVC AGB KIT: 3-L 12 FR X 16 CM/CDC-22123-XCN1A, LBCVC AGB KIT: 3-L 12 FR X 20 CM/CDC-25123-XCN1A, LBCVC AGB KIT:3-L 12 FR X 20 CM/CDC-25123-XN1A
- Reason / hazard
- Lidocaine, Bupivacaine, and 0.9% sodium chloride, the subject of a supplier recall, were included into kits and sets. The supplier was found to have quality issues and limited reports of drug ineffectiveness, and if sterility of a water-based injectable product is compromised this could lead to serious or life-threatening injuries.
- Classification
- Class I
- Status
- Ongoing
- Distribution / sold at
- US Nationwide distribution including in the states of VA, NY, UT, SC, LA, MN, PA, AZ, CA, MI, OH, FL, CT, KS, IL, TX, WI, GA, NC, MD, OR, AK, NH, IA, ND, OK, NM, AL, MA, KY, WA, AR, CO, IN, ME, TN, WY, NJ, SD, MS, MO, MT, DE, ID, NE, WV, RI, VT, NV, HI, DC, PR.
- Recall initiated
- 2026-04-21
- Code info
- REF(Material)/UDI-DI/Lot(Batch): AK-15122-F/10801902197136, 10801902193275, 10801902168525/33F24E0348, 33F24J0280, 33F24K0119, 33F24L0601, 33F25A0477, 33F25C0908, 33F24A0850; ASK-12123-UOC1/10801902193275/33F25B0430, 33F25D0119; ASK-12123-UPM/10801902193275, 10801902222463/33F25E0535, 33F25G0188, 33F25G0189, 33F26C0333, 33F25D0262; ASK-15123-UOC1/10801902193275/33F25C0378, 33F25D0121; ASK-15123-UPM/10801902193275, 10801902159950/33F25E0718, 33F25H0447, 33F26A1389, 33F24E0649; CDC-12123-P1A/40801902111768, 10801902193275/33F24D0632, 33F24F0758, 33F24L0465, 33F25A0281, 33F25E0794; CDC-12123-XP1A/10801902193275, 30801902111785, 10801902213416/33F25J0723, 33F26A1382, 33F26B0433, 36F25A0007, 36F25B0009, 36F25E0018, 33F24D0299; CDC-12123-XPCN1A/10801902213416, 10801902193275/33F24D0950, 33F24K0417, 33F24L0466, 33F25A0216, 33F25B0667, 33F25D0436, 33F25D0666, 33F25E0693, 33F25G0239, 33F25J0247, 33F26A1556, 33F26B0801, 33F26C0767, 33F25B0668; CDC-12123-XPN1A/10801902193275, 10801902213409/33F25D0554, 33F25G0240, 33F25J0312, 33F26B0214,33F26C0199, 33F25A0181; CDC-15123-P1A/10801902193275/33F25D0437, 33F25H0547,33F25B0613; CDC-15123-XPCN1A/10801902213430, 10801902193275/33F24F0107, 33F24F0998, 33F24K0403, 33F24L0834, 33F25A0283, 33F25B0527, 33F25E0650, 33F25E0745, 33F25F0085, 33F25H0596, 33F25J0314, 33F26A1464, 33F26B0883, 33F25B0669; CDC-15123-XPN1A/0801902193275, 10801902213423, 10801902116793/33F25D0439, 33F25G0238, 33F25H0597, 33F25J0315, 33F26A1468, 33F26B0826, 33F24C0461; CDC-16123-XPCN1A/10801902193275/33F25B0671, 33F25D0440, 33F25E0651, 33F25G0236, 33F25H0529, 33F25J0316, 33F25G0237, AK-25122-CDC/10801902193275/33F24G0370, 33F24K0135, 33F25C0120, 33F25C0969, 33F25E0942, 33F24G0371; AK-25142-F/10801902193275, 70801902123205, 10801902167795/33F25C0863, 33F25E0087, 33F25H0038, 33F25J0846, 33F26B0531, 33F24D0089; ASK-22122-MM1/10801902193275/33F25E0546; ASK-22122-MM2/10801902193275, 10801902221923, 10801902144956/33F25G0172, 33F25L0031, 33F25M0310, 33F26A1652, 33F26B0151, 33F26B0450, 33F24F0854; ASK-22142-UOC1/10801902193275/33F25C0545, 33F25D0122; ASK-22142-UPM/10801902193275, 10801902222487, 10801902202281/33F25D0594, 33F25F0116, 33F25G0187, 33F26A1390, 33F26B0205, 33F24D0396; ASK-25142-UNC1/10801902193275/33F25C0655, 33F24K0561; ASK-25142-UOC1/10801902193275/33F25B0362, 33F25D0123; ASK-25142-UPM/10801902193275/33F25E0595, 33F25H0446, 33F25J0381, 33F26A1446, 33F25J0092; ASK-26142-UNC1/10801902193275/33F25B0427, 33F25K0058; AU-22122-F/10801902193275/33F24K0269, 33F25A0584, 33F25C0862, 33F25E0245, 33F25J0567; CDC-22122-1A/20801902123484, 10801902193275/, CDC-22122-XCN1A-U/33F24F0996, 33F24L0374, 33F25C0461; CDC-22122-XN1A/10801902193275/33F25C0239, 33F25E0875, 33F25J0419; CDC-22122-XN1A-U/10801902193275, 10801902213621/33F25A0309, 33F25E0876, 33F26B0217; CDC-22142-1A/50801902123515, 10801902193275/33F24D0928, 33F24F0103, 33F24F0896, 33F24L0111, 33F24L0955, 33F25B0500, 33F25D0447, 33F25E0654, 33F25G0278, 33F25H0577, 33F25J0319; CDC-22142-XC1A/10801902193275/33F25A0284, 33F25J0180; CDC-22142-XCN1A/10801902193275, 10801902213539/33F25A0285, 33F25B0529, 33F25D0449, 33F25E0655, 33F25E0818, 33F25F0214, 33F25G0233, 33F26A1477, 33F26B0898, 33R25J0455; CDC-22142-XCN1A-C/10801902193275/33F24L0473, 33F25B0615, 33F25F0213; CDC-22142-XN1A/10801902193275, 10801902213522/33F25C0240, 33F25E0578, 33F25F0215, 33F25H0528, 33F25J0321, 33F26A1383, 33F26B0215, 33F26C0016; CDC-22142-XN1A-C/10801902193275/33F25B0485, 33F25E0879; CDC-23122-XCN1A-U/10801902193275/33F24L0279, 33F25D0492; CDC-23122-XN1A-U/10801902193275/33F25E0579; CDC-23122-XU1A/10801902193275/33F25D0493, 33F25H0432, 33F25J0248; CDC-25122-1A/10801902193275/33F25B0619, 33F25E0881; CDC-25122-X1A/10801902193275/33F25E0882, 33F25J0422; CDC-25122-XCN1A/10801902213492, 10801902193275/33F24F0929, 33F25C0242, 33F25D0498, 33F25E0883, 33F25G0234, 33F26B0439, 33F26C0214; CDC-25122-XCN1A-U/10801902193275/33F24L1041, 33F25B0620, 33F25F0121, 33F25G0226; CDC-25122-XN1A/10801902193275, 10801902213485/33F25C0243, 33F25E0884, 33F25K0183, 33R25E0884; CDC-25122-XU1A/10801902193275/33F25B0621, 33F25E0886, 33F25H0568; CDC-25142-1A/50801902123584, 10801902193275/33F24F0837, 33F24K0457, 33F24L0113, 33F24L1031, 33F25B0622, 33F25D0503, 33F25E0889, 33F25F0106, 33F25J0127; CDC-25142-XC1A/10801902193275/33F25B0467; CDC-25142-XCN1A/10801902193275, 10801902213553/33F25A0312, 33F25E0657, 33F25E0821, 33F25G0231, 33F25H0527, 33F25J0428, 33F26B0452, 33F26C0133; CDC-25142-XCN1A-C/10801902193275/33F24L0377, 33F25D0238; CDC-25142-XN1A/10801902193275, 10801902213546/33F25B0624, 33F25D0504, 33F25E0823, 33F25G0232, 33F26B0437; CDC-25142-XN1A-C/10801902193275/33F24L0485, 33F25B0676; CDC-26122-1A/60801902123611, 10801902193275/33F24F0084, 33F24K0475, 33F25G0389; CDC-26122-XCN1A/10801902213515, 10801902193275/33F24F1028, 33F24L0117, 33F25D0308, 33F25D0557, 33F25F0219; CDC-26122-XN1A/10801902193275/33F25E0451; CDC-26142-1A/10801902193275/33F24L0488, 33F24L1025, 33F25C0368, 33F25E0891, 33F25H0406, 33F25J0116; CDC-26142-X1A/10801902193275/33F25D0558, 33F25E0892; CDC-26142-XCN1A/10801902193275, 10801902213577/33F24L0941, 33F25B0530, 33F25D0507, 33F25E0699, 33F25H0525, 33F26B0396, 33F26B0895, 33F26C0780; CDC-26142-XN1A/10801902193275, 10801902213560/33F24L0927, 33F25E0580, 33F25H0526, 33F26B0393; AK-25122-F/ CDC-22122-XCN1A/10801902193275/33F24G0371, 33F24L0742, 33F25C0822, 33F25E0084, 33F25J0483, 33F25L0400; CDC-22123-XCN1A/10801902213591, 10801902193275/33F24D0842,33F24F1018, 33F24L0110, 33F25C0205, 33F25D0613, 33F25E0878, 33F26B0455; CDC-25123-XCN1A/10801902213614, 10801902193275/33F24D0818, 33F24L0483, 33F25C0244, 33F25D0502, 33F25E0888, 33F25G0228, 33F25J0425, 33F26B0395, 33F26B0894; CDC-25123-XN1A/10801902193275/33F25C0332, 33F25G0229
- Product quantity
- 62,719
- Voluntary mandated
- Voluntary: Firm initiated
- City
- Morrisville
- State
- NC
- Country
- United States
- Event id
- 98793