Fresenius Kabi USA, LLC: Ivenix Infusion System (IIS), Large Volume Pump, Model Number: LVP-0004
Read the official notice at FDA (device) →
The text below is copied from the official FDA (device) record. If anything differs, the official notice is correct.
- Company / brand
- Fresenius Kabi USA, LLC
- Product
- Ivenix Infusion System (IIS), Large Volume Pump, Model Number: LVP-0004
- Reason / hazard
- Fresenius Kabi has been made aware via customer complaints that dropping or severe jarring of the Ivenix LVP may cause damage to internal components that may not be apparent or readily visible and the devices therefore remained in service. More specifically, drops or jarring on a hard surface can cause the touchscreen input cable to loosen, resulting in unintended touchscreen behavior on the LVP, including random screen touches and/or lack of screen responsiveness to user input. Under the reported failure condition, unintended touch inputs may register at arbitrary screen locations without user interaction. These inputs occur intermittently and without a predictable pattern.
- Classification
- Class I
- Status
- Ongoing
- Distribution / sold at
- Domestic: CA, CO, FL, GA, ID, IL, MD, MI, MI , MN, MO, MS, NE, NJ, NV, OK, OR, SC, TX, VA, WA, WI, WI;
- Recall initiated
- 2026-05-06
- Code info
- Model Number: LVP-0004; UDI-DI: 00811505030320; All Serial/Lot Numbers:
- Product quantity
- 18,444
- Voluntary mandated
- Voluntary: Firm initiated
- City
- North Andover
- State
- MA
- Country
- United States
- Event id
- 98945