Siemens Medical Solutions USA, Inc: ARTIS Icono biplane, floor and ceiling.: ARTIS icono biplane Model 11327600. ARTIS icono ceiling Model 11328100. ARTIS icono floor Model 113
Read the official notice at FDA (device) →
Get alerts for Siemens Medical Solutions
The text below is copied from the official FDA (device) record. If anything differs, the official notice is correct.
- Company / brand
- Siemens Medical Solutions USA, Inc
- Product
- ARTIS Icono biplane, floor and ceiling.: ARTIS icono biplane Model 11327600. ARTIS icono ceiling Model 11328100. ARTIS icono floor Model 11327700. interventional fluoroscopic x-ray system
- Reason / hazard
- During patient examinations, the system may sporadically display a 0 x-ray dose administered to the patient. This behavior has only been observed during fluoroscopy and/or single-frame acquisitions. Despite the reported x-ray dose showing 0, an actual x-ray dose is applied to the patient.
- Classification
- Class II
- Status
- Ongoing
- Distribution / sold at
- US and Worldwide
- Recall initiated
- 2026-03-09
- Code info
- ARTIS icono biplane UDI 4056869063317 ARTIS icono ceiling UDI 4056869295923 ARTIS icono floor UDI 4056869149325
- Voluntary mandated
- FDA Mandated
- City
- Malvern
- State
- PA
- Country
- United States
- Event id
- 98992