MICROVENTION INC.: LVIS Intraluminal Support Device, REF/Outside Diameter(OD)xTotal Length(TL)xWorking Length(WL): 214022-CAS/4.0 mm x 22 mm x 18 mm, 214518-CA
Read the official notice at FDA (device) →
The text below is copied from the official FDA (device) record. If anything differs, the official notice is correct.
- Company / brand
- MICROVENTION INC.
- Product
- LVIS Intraluminal Support Device, REF/Outside Diameter(OD)xTotal Length(TL)xWorking Length(WL): 214022-CAS/4.0 mm x 22 mm x 18 mm, 214518-CAS/4.5 mm x 18 mm x 14 mm, 213517-CAS/3.5 mm x 17 mm x 13 mm
- Reason / hazard
- Stent-assisted coiling device may due to commingling during manufacturing have dimensional nonconformance and/or incorrect device configuration. If device is oversized may lead to increased vessel wall stress, neointimal hyperplasia, vessel injury/rupture; if undersized may lead to incomplete expansion/malposition, acute stent thrombosis, distal embolization, ischemic neurological events
- Classification
- Class II
- Status
- Ongoing
- Distribution / sold at
- International distribution to the country of China.
- Recall initiated
- 2026-04-06
- Code info
- Lots: 0000616462, 0000619851, 0000552987, 0000551725. UDI: (01)00842429115893(11)240325(17)270228(10)0000552987, (01)00842429115893(11)240325(17)270228(10)0000552987, (01)00842429115893(11)240325(17)270228(10)0000552987, (01)00842429115930(11)240325(17)270228(10)0000551725, (01)00842429115930(11)240325(17)270228(10)0000551725, (01)00842429115930(11)240610(17)270531(10)0000616462, (01)00842429115930(11)240610(17)270531(10)0000616462, (01)00842429115961(11)240610(17)270531(10)0000619851, (01)00842429115961(11)240610(17)270531(10)0000619851, (01)00842429115961(11)240610(17)270531(10)0000619851, (01)00842429115961(11)240610(17)270531(10)0000619851,
- Product quantity
- 38
- Voluntary mandated
- Voluntary: Firm initiated
- City
- Aliso Viejo
- State
- CA
- Country
- United States
- Event id
- 98755
Free daily digest + "My products" watchlist