Heraeus Medical GmbH (Dental Division): Heraeus, PALAMIX duo. Material Number: 66057897.
Read the official notice at FDA (device) →
Get alerts for Heraeus Medical GmbH
The text below is copied from the official FDA (device) record. If anything differs, the official notice is correct.
- Company / brand
- Heraeus Medical GmbH (Dental Division)
- Product
- Heraeus, PALAMIX duo. Material Number: 66057897.
- Reason / hazard
- Based on long-term stability testing, it was determined that in units older than 1.5 years (17 months) the mixing rod may, in rare cases, come loose from its lock during the mixing process. Notification is to reduce the shelf life from 3 years to 1.5 years (17 months).
- Classification
- Class III
- Status
- Ongoing
- Distribution / sold at
- Worldwide distribution - US Nationwide and the countries of Aland Islands, Algeria, Australia, Austria, Belgium, Brazil , Chile, China, Cyprus, Czech Republic, Denmark, Ecuador, Finland, France, Germany, Greece, Hungary, India, Ireland, Israel, Italy, Kosovo, Lithuania, Luxembourg, Malaysia, Malta, Mexico, Netherlands, Nigeria, Norway, Peru, Philippines, Poland, Portugal, Puerto Rico, Rep. of Korea, Saudi Arabia, Singapore, Slovakia, Slovenia, South Africa, Spain, Sweden, Switzerland, Thailand, United Kingdom, Uruguay, Utd.Arab Emir.;
- Recall initiated
- 2026-03-25
- Code info
- Material Number: 66057897; UDI-DI: 04260102133617; Batch: all batches with numbers below 520215;
- Product quantity
- 1,125 units
- Voluntary mandated
- Voluntary: Firm initiated
- City
- Wehrheim
- Country
- Germany
- Event id
- 98608