Remel Europe Ltd.: Oxoid Agglutinating Sera, Salmonella 9-0 R30957301

Medical DevicesSource: FDA (device) · Date: 2026-04-29

Read the official notice at FDA (device) →

Share on XFacebookRedditEmail

Guides: Which agency handles which recall (FDA vs USDA vs CPSC vs NHTSA)

The text below is copied from the official FDA (device) record. If anything differs, the official notice is correct.

Company / brand
Remel Europe Ltd.
Product
Oxoid Agglutinating Sera, Salmonella 9-0 R30957301
Reason / hazard
Affected lots are not performing to IFU criteria showing cross reactions with 2-0 within 60 seconds. Cross-reactivity may result in the misidentification of a Salmonella 9-O serotype as a 2-O serotype.
Classification
Class II
Status
Ongoing
Distribution / sold at
Worldwide - US Nationwide distribution to the states of AL, CA, FL, IA, IL, IN, MI, MN, MO, NC, NE, OH, SD, TX, WA.
Recall initiated
2026-03-25
Code info
UDI-DI 05056080500270 lots 6172600, 6172741, 6171393, 6150168, 6107767, 3766399, 3766400, 3780497, 3719592, 3709358, 6266130, 6253410, 6260215, 6222350, 6233859, 6233858
Product quantity
117 US; 950 OUS
Voluntary mandated
Voluntary: Firm initiated
City
Dartford
State
N/A
Country
United Kingdom
Event id
98652
Free daily digest + "My products" watchlist

Get the daily digest, and tell us what to watch. Everything is optional except your email.

Free allergen alerts:

Categories:

Privacy: we store only what you enter here, use it only to send recall alerts you asked for, never sell or share it, and you can unsubscribe anytime. Double opt-in: we will email you a confirmation link before sending anything; alerts start only after you confirm. Every email has unsubscribe and manage-watchlist links.