Siemens Medical Solutions USA, Inc: ARTIS Pheno VE30A and VE40A, Model 10849000
Read the official notice at FDA (device) →
Get alerts for Siemens Medical Solutions
The text below is copied from the official FDA (device) record. If anything differs, the official notice is correct.
- Company / brand
- Siemens Medical Solutions USA, Inc
- Product
- ARTIS Pheno VE30A and VE40A, Model 10849000
- Reason / hazard
- During 3D acquisitions, lighter and darker patient images may be captured, which may result in less accurate 3D reconstruction. The variation of brightness is a result of a constant unregulated medium dose and the angular change of patient diameter caused by rotational acquisition. This may result in unintentional low-dose radiation exposure to a patient.
- Classification
- Class II
- Status
- Ongoing
- Distribution / sold at
- U.S. and O.U.S.
- Recall initiated
- 2026-03-12
- Code info
- UDI 04056869046877
- Product quantity
- 73
- Voluntary mandated
- FDA Mandated
- City
- Malvern
- State
- PA
- Country
- United States
- Event id
- 98772
Free daily digest + "My products" watchlist