Cook Incorporated: COOK MEDICAL Cook Staged Extubation Set: Reference Part Number C-CAE-14.0-83-EXTU, Order Number G24073
Read the official notice at FDA (device) →
The text below is copied from the official FDA (device) record. If anything differs, the official notice is correct.
- Company / brand
- Cook Incorporated
- Product
- COOK MEDICAL Cook Staged Extubation Set: Reference Part Number C-CAE-14.0-83-EXTU, Order Number G24073
- Reason / hazard
- Products from the affected device lots were labeled with expiration dates that exceed the true shelf life.
- Classification
- Class II
- Status
- Ongoing
- Distribution / sold at
- Worldwide distribution.
- Recall initiated
- 2026-03-05
- Code info
- Reference Part Number C-CAE-14.0-83-EXTU, UDI (01)00827002240735(17)270509(10)16010654, Lot Number 16010654.
- Product quantity
- 95 units
- Voluntary mandated
- Voluntary: Firm initiated
- City
- Bloomington
- State
- IN
- Country
- United States
- Event id
- 98495
Free daily digest + "My products" watchlist