Ion Beam Applications S.A.: IBA Proton Therapy System - PROTEUS 235
Read the official notice at FDA (device) →
The text below is copied from the official FDA (device) record. If anything differs, the official notice is correct.
- Company / brand
- Ion Beam Applications S.A.
- Product
- IBA Proton Therapy System - PROTEUS 235
- Reason / hazard
- It was identified that eh Universal Beam Triggering Interface (UBTI) is disabled in the Therapy Safety System (TSS) configuration. In this condition, the TSS does not interrupt the beam delivery if the Healthy Signal from an external gating device is lost. This may lead to a risk of mistreatment when external gating devices are used.
- Classification
- Class II
- Status
- Ongoing
- Distribution / sold at
- Worldwide - US Nationwide distribution in the state of FL and the countries of Italy and Japan.
- Recall initiated
- 2026-02-03
- Code info
- PROTEUS 235; UDI-DI: 05404013801138; Serial Numbers: SBF103 (JP), SBF124 (EU), SBF133 (US), SBF140 (EU)
- Product quantity
- 4 units (1 US, 3 OUS)
- Voluntary mandated
- Voluntary: Firm initiated
- City
- Ottignies-Louvain-La-Neuve
- State
- N/A
- Country
- Belgium
- Event id
- 98364