Siemens Healthcare Diagnostics, Inc.: Atellica CH Urine Albumin (UAlb). Material Number: 11537225
Read the official notice at FDA (device) →
Get alerts for Siemens Healthcare Diagnostics
The text below is copied from the official FDA (device) record. If anything differs, the official notice is correct.
- Company / brand
- Siemens Healthcare Diagnostics, Inc.
- Product
- Atellica CH Urine Albumin (UAlb). Material Number: 11537225
- Reason / hazard
- Falsely depressed UAlb patient results may occur. Affected samples with any urine albumin concentration above the UAlb measuring interval >40.0 mg/dL may report values between 6.2 mg/dL and 40.0 mg/dL.
- Classification
- Class II
- Status
- Ongoing
- Distribution / sold at
- Worldwide distribution - US Nationwide and the countries of Australia, Austria, Belgium, Brazil, Canada, Chile, Curacao,¿St¿Eus, Czech¿Republic, Denmark, Finland, France, Germany, Greece, Hong Kong, Ireland, Israel, Italy, Korea, Latvia, Malaysia, Netherlands, New Zealand, Norway, Portugal, Romania, Singapore, Slovakia, South Africa, Spain, Sweden, Switzerland, United¿Kingdom.
- Recall initiated
- 2026-01-20
- Code info
- Material Number: 11537225. UDI-DI: 00630414611099. Lot Numbers: All lot numbers.
- Product quantity
- 4,885 units
- Voluntary mandated
- Voluntary: Firm initiated
- City
- Tarrytown
- State
- NY
- Country
- United States
- Event id
- 98370
Free daily digest + "My products" watchlist