Canon Medical System, USA, INC.: Alphenix INFX-8000H, interventional fluoroscopic x-ray system
Read the official notice at FDA (device) →
Get alerts for Canon Medical System
The text below is copied from the official FDA (device) record. If anything differs, the official notice is correct.
- Company / brand
- Canon Medical System, USA, INC.
- Product
- Alphenix INFX-8000H, interventional fluoroscopic x-ray system
- Reason / hazard
- It has been discovered that there is a possibility that the fixing screws of the ceiling movement gear in the ceiling suspended arm that are part of the X-ray interventional systems may become loose. As a result, ceiling lateral movement may no longer be possible, abnormal noise may be generated from the suspension system and a sensor error may be displayed
- Classification
- Class II
- Status
- Ongoing
- Distribution / sold at
- Worldwide distribution - US Nationwide and the country of Dominican Republic.
- Recall initiated
- 2025-12-19
- Code info
- serial numbers: D2A1672385, A2A20Z2165, B2A1692426.
- Product quantity
- 3 units
- Voluntary mandated
- Voluntary: Firm initiated
- City
- Tustin
- State
- CA
- Country
- United States
- Event id
- 97970
Free daily digest + "My products" watchlist