Siemens Medical Solutions USA, Inc: Interventional Fluoroscopic X-Ray System: LUMINOS Q.namix R; Siemens Material Number (SMN): 11574002; With Software Version VB10D-SP02
Read the official notice at FDA (device) →
Get alerts for Siemens Medical Solutions
The text below is copied from the official FDA (device) record. If anything differs, the official notice is correct.
- Company / brand
- Siemens Medical Solutions USA, Inc
- Product
- Interventional Fluoroscopic X-Ray System: LUMINOS Q.namix R; Siemens Material Number (SMN): 11574002; With Software Version VB10D-SP02
- Reason / hazard
- Ortho images acquired with preset automatic horizontal flip are not flipped and when acquiring single images, there is the possibility that an image of a previous patient may be processed.
- Classification
- Class II
- Status
- Ongoing
- Distribution / sold at
- Worldwide - US Nationwide distribution in the states of AK, CA, IL, MN, NY, OR, VA, WI and the countries of Austria, Belgium, Denmark, France, Germany, Greece, Ireland, Italy, Netherlands, Norway, South Africa, Spain, Sweden, Switzerland.
- Recall initiated
- 2025-12-29
- Code info
- Siemens Material Number (SMN): 11574002; UDI-DI: 04056869993973; Serial Numbers: 150050, 150035, 150033, 150034, 150048, 150021, 150020, 150053, 150023, 150044, 150049, 150037, 150054, 150011, 150059, 150024, 150038, 150051, 150029, 150045, 150032, 150056, 150043, 150055, 150041, 150026, 150025, 150022, 150031, 150039, 150046, 150042, 150028, 150047, 150030, 150052, 150027, 150057, 150040, 150058, 150063;
- Product quantity
- 41 units
- Voluntary mandated
- Voluntary: Firm initiated
- City
- Malvern
- State
- PA
- Country
- United States
- Event id
- 98244