⚠ RecallRadar📷 ScanGuidesOfficial U.S. recall notices, restated plainly

Sysmex America, Inc.: TS-10/TS-10H Tube Sorter, automatic sorting system for sorting barcoded sample tubes, Catalog Number BN168792, running software versions 00-

Medical DevicesSource: FDA (device) · Date: 2026-01-28

Read the official notice at FDA (device) →

Share on XFacebookRedditEmail

Guides: Which agency handles which recall (FDA vs USDA vs CPSC vs NHTSA)

The text below is copied from the official FDA (device) record. If anything differs, the official notice is correct.

Company / brand
Sysmex America, Inc.
Product
TS-10/TS-10H Tube Sorter, automatic sorting system for sorting barcoded sample tubes, Catalog Number BN168792, running software versions 00-17
Reason / hazard
Under specific conditions, the lift's electromagnetic holding force may disengage unexpectedly, allowing unintended downward movement of the lift.
Classification
Class II
Status
Ongoing
Distribution / sold at
US Nationwide.
Recall initiated
2025-12-15
Code info
Serial Numbers: 11281 11611 11918 11790 12040 11991 12218 11950 11795 11533 11635 12054 11610 11453 11753 11539 11990 12298 11774 11541 11981 11788 11834 12005 11907 11524 12237 11246 12154 12104 12299 12176 11711 11420 12153 12177 12074 11405 11303 12039 11250 11336 12137 12075
Product quantity
44 units
Voluntary mandated
Voluntary: Firm initiated
City
Lincolnshire
State
IL
Country
United States
Event id
98213
Free daily digest + "My products" watchlist

Get the daily digest, and tell us what to watch. Everything is optional except your email.

Free allergen alerts:

Categories:

Privacy: we store only what you enter here, use it only to send recall alerts you asked for, never sell or share it, and you can unsubscribe anytime. Double opt-in: we will email you a confirmation link before sending anything; alerts start only after you confirm. Every email has unsubscribe and manage-watchlist links.