Beckman Coulter Ireland, Inc.: The LDL-Cholesterol assay is comprised of two distinct phases. In phase one a unique detergent solubilizes cholesterol from non-LDL- lipopro

Medical DevicesSource: FDA (device) · Date: 2026-01-28

Read the official notice at FDA (device) →

Share on XFacebookRedditEmail

Guides: Which agency handles which recall (FDA vs USDA vs CPSC vs NHTSA)

The text below is copied from the official FDA (device) record. If anything differs, the official notice is correct.

Company / brand
Beckman Coulter Ireland, Inc.
Product
The LDL-Cholesterol assay is comprised of two distinct phases. In phase one a unique detergent solubilizes cholesterol from non-LDL- lipoprotein particles. This cholesterol is consumed by cholesterol esterase, cholesterol oxidase, peroxidase and 4- aminoantipyrine to generate a colorless end product. In phase two a second detergent in reagent 2 releases cholesterol from the LDL lipoproteins. This cholesterol reacts with cholesterol esterase, cholesterol oxidase and a chromogen system to yield a blue color complex which can be measured bichromatically at 540/660nm. The resulting increase in absorbance is directly proportional to the LDL-C concentration in the sample.
Reason / hazard
Beckman Coulter identified that LDL Cholesterol OSR6x96 is not meeting the performance claim listed in the Instructions for Use (IFU). LDL Cholesterol reagent showed clinically significant interference due to triglyceride concentration up to 1500 mg/dL may cause a maximum positive bias of up to 28.8% in low LDL Cholesterol samples (concentration 80 mg/dL) and up to 11.02% in high LDL Cholesterol samples (concentration 160mg/dL).
Classification
Class II
Status
Ongoing
Distribution / sold at
US: MD, NJ, OH, FL, CA, TX, NC, WA, MA, NV, GA, CO, PA, NM, UT, VA, KS, MT, OK, NY, AZ, LA, OR, AL, MO, IN, IL, KY, MS, NH, AK, SD, MN, TN, PR, MI, IA, WI, WV, ME, SC, HI, RI, ND, NE, ID, AR, WY, DC, GU, DE, CT. OUS: Brazil, South Africa, Argentina, Mexico, Canada, Costa Rica, Peru, El Salvador, Ecuador, Panama, Singapore, Ireland, United Kingdom of Great Britain and Northern Ireland, Germany, China, Korea, Republic of
Recall initiated
2025-12-08
Code info
REF/UDI-DI: OSR6196/15099590010928, OSR6296/15099590020361. All Lots and Expiration Dates. IFU Version BAOSR6X96-10 and prior.
Product quantity
34,451
Voluntary mandated
Voluntary: Firm initiated
City
O'Callaghan'S Mills
Country
Ireland
Event id
98094
Free daily digest + "My products" watchlist

Get the daily digest, and tell us what to watch. Everything is optional except your email.

Free allergen alerts:

Categories:

Privacy: we store only what you enter here, use it only to send recall alerts you asked for, never sell or share it, and you can unsubscribe anytime. Double opt-in: we will email you a confirmation link before sending anything; alerts start only after you confirm. Every email has unsubscribe and manage-watchlist links.