Canon Medical System, USA, INC.: VANTAGE TITAN 3T Model MRT-3010/MEXL-3010

Medical DevicesSource: FDA (device) · Date: 2026-01-21

Read the official notice at FDA (device) →

Share on XFacebookRedditEmail

Get alerts for Canon Medical System

Guides: Which agency handles which recall (FDA vs USDA vs CPSC vs NHTSA)

The text below is copied from the official FDA (device) record. If anything differs, the official notice is correct.

Company / brand
Canon Medical System, USA, INC.
Product
VANTAGE TITAN 3T Model MRT-3010/MEXL-3010
Reason / hazard
There is a potential for formation of ice occurring in the venting system of the superconducting magnet of the MRI systems. if a quench occurs in this condition, helium gas may not be able to escape via the designed vent paths, and the pressure in the helium vessel may rise. This pressure build-up could eventually rupture the helium vessel, causing helium gas to be released into the MR scanning room.
Classification
Class II
Status
Ongoing
Distribution / sold at
U.S. Nationwide distribution in the states of CA, CO, FL, IA, IL, IN, KS, MD, MO, MT, NV, NY, OH, OR, PA, PR, TX, VA, and WI.
Recall initiated
2025-12-19
Code info
Model Number: MRT-3010/MEXL-3010 UDI-DI code: 04987670101649 CMSC Model Number/Model Number/Serial Numbers/SID/UDI code: MRT-3010/A5 MEXL-3010/A5 A5D14X2031 30000041 MRT-3010/A5 MEXL-3010/A5 A5D13Y2025 30001708 MRT-3010/A5 MEXL-3010/A5 A5D13Y2026 30005894 MRT-3010/A5 MEXL-3010/A5 A5G1692039 30007620 (01)04987670101649(21)A5G1692039 MRT-3010/A5 MEXL-3010/A5.007 A5B11Y2006 30019175 MRT-3010/A7 MEXL-3010/S7 S7A1642006 30024266 (01)04987670101649(21)A7A1642001 MRT-3010/A5 MEXL-3010/A5 A5G1692038 30027318 (01)04987670101649(21)A5G1692038 MRT-3010/A5 MEXL-3010/A5 A5G1642036 30029482 (01)04987670101649(21)A5G1642036 MRT-3010/A5 MEXL-3010/A5.007 A5F1562034 30050314 MRT-3010/A5 MEXL-3010/A5 A5B11Z2007 303354 MRT-3010/A5 MEXL-3010/A5 A5C1252012 336719 MRT-3010/A5 MEXL-3010/A5 A5D1312019 350553 MRT-3010/A5 MEXL-3010/A5 A5B1192003 350579 MRT-3010/A5 MEXL-3010/A5 A5C12X2016 350918 MRT-3010/A5 MEXL-3010/A5 A5C12Y2017 351122 MRT-3010/A5 MEXL-3010/A5 A5C1212009 351387 MRT-3010/A5 MEXL-3010/A5 A5B11Y2005 351452 MRT-3010/A5 MEXL-3010/A5 A5C1262013 356865 MRT-3010/A5 MEXL-3010/A5 A5C1292015 378015 MRT-3010/A5 MEXL-3010/A5 A5C1412028 383946 MRT-3010/A5 MEXL-3010/A5 A5E1522033 389630 MRT-3010/A5 MEXL-3010/A5 A5D1362021 390397
Product quantity
22 systems
Voluntary mandated
Voluntary: Firm initiated
City
Tustin
State
CA
Country
United States
Event id
98112
Free daily digest + "My products" watchlist

Get the daily digest, and tell us what to watch. Everything is optional except your email.

Free allergen alerts:

Categories:

Privacy: we store only what you enter here, use it only to send recall alerts you asked for, never sell or share it, and you can unsubscribe anytime. Double opt-in: we will email you a confirmation link before sending anything; alerts start only after you confirm. Every email has unsubscribe and manage-watchlist links.