Medline Industries, LP: MEDLINE beds labeled as: Full Electric Low Lightweight Homecare Bed, REF MDR107003LO
Read the official notice at FDA (device) →
Get alerts for Medline Industries
The text below is copied from the official FDA (device) record. If anything differs, the official notice is correct.
- Company / brand
- Medline Industries, LP
- Product
- MEDLINE beds labeled as: Full Electric Low Lightweight Homecare Bed, REF MDR107003LO
- Reason / hazard
- Medline Industries has identified a potential hazard involving the hand control pendant and associated wires. In rare and atypical scenarios such as operating the bed above its specified weight limit, encountering an obstruction to bed movement, or experiencing jammed components/motor the hand pendant and pendant cords may overheat and, in rare instances, pose a risk of fire. Medline Industries, LP has received a total of (58) complaints associated with reports of pendant sparking, burning, melting, smoking or fire.
- Classification
- Class I
- Status
- Ongoing
- Distribution / sold at
- US, CA, PA, GUAM, VI
- Recall initiated
- 2025-11-26
- Code info
- UDI/DI 40080196325901, All lots
- Product quantity
- 15581 units
- Voluntary mandated
- Voluntary: Firm initiated
- City
- Northfield
- State
- IL
- Country
- United States
- Event id
- 98185
Free daily digest + "My products" watchlist