Beaver-Visitec International, Inc.: BVI Leos Laser and Endoscopy System. Model Number: OME6000U.
Read the official notice at FDA (device) →
The text below is copied from the official FDA (device) record. If anything differs, the official notice is correct.
- Company / brand
- Beaver-Visitec International, Inc.
- Product
- BVI Leos Laser and Endoscopy System. Model Number: OME6000U.
- Reason / hazard
- Manufacturing deviation in which the ferrites specified for the USB cable assembly linking the Nurse Monitor touchscreen to the System Controller were not installed as intended. The omission of the ferrites on the USB communication cable results in non-conformance to the design specifications, which require two ferrites (one at each end of the cable).
- Classification
- Class II
- Status
- Ongoing
- Distribution / sold at
- US Nationwide distribution in the states of FL, MI, OK, TX, WI.
- Recall initiated
- 2025-12-05
- Code info
- Model Number: OME6000U. UDI: (01)00886158037576(11)250513(21). Serial Numbers: 77UBV500003, 77UBV500004, 77UBV500005, 77UBV500006, 77UBV500010.
- Product quantity
- 5 units
- Voluntary mandated
- Voluntary: Firm initiated
- City
- Waltham
- State
- MA
- Country
- United States
- Event id
- 98076