CareFusion 303, Inc.: BD Pyxis MedStation ES, REF: 323, BD Pyxis MedStation ES 7 Drawer Auxiliary, REF: 324, BD Pyxis MedStation ES Tower, REF: 352, BD Pyx
Read the official notice at FDA (device) →
The text below is copied from the official FDA (device) record. If anything differs, the official notice is correct.
- Company / brand
- CareFusion 303, Inc.
- Product
- BD Pyxis MedStation ES, REF: 323, BD Pyxis MedStation ES 7 Drawer Auxiliary, REF: 324, BD Pyxis MedStation ES Tower, REF: 352, BD Pyxis Pro 7-Drawer Auxiliary, REF: 1149-00, BD Pyxis Pro 7- Drawer Auxiliary with Secure Tower, REF: 1152-00, BD Pyxis Pro MedStation Main, REF: 1155-00, BD Pyxis CII Safe ES, REF: 1116-00
- Reason / hazard
- Automated Dispensing Cabinets (ADC) received a Half Height CUBIE drawer firmware update, which caused a Cubie Insert event which prompts software to generate a duplicate address, which causes an error and the drawer to fail, which may lead to inability or delay in accessing stored items, delay in the replenishment of ADCs or patient specific medications stored outside of ADCs.
- Classification
- Class II
- Status
- Ongoing
- Distribution / sold at
- Worldwide - US Nationwide distribution including in the states of CO, IN, OH, TX, KS, FL, GA, NC, CA, HI, IL, NY, MI, VA, AL, LA, NE, KY, OK, AR, TN, WY, NJ, WI, SD, AZ, NV, MO, MA, ID, WV, DE, PA, WA, MT, CT, IA, MD, PR, MN, NM, OR, VT, MS, ME, AK, UT, DC, GU, SC, NH, RI, ND, AS and the countries of CH, CA, KW, QA, NL, GB, UY, AU, NZ, BH, TH, SG, CN, FR, LB, BS, HK, TR, PL, AR, BR, CL, CO, MX, ES, BM, SA, IL.
- Recall initiated
- 2025-11-20
- Code info
- Firmware Version: 1.48. REF/UDI-DI: 323/10885403512667, 324/10885403512681, 352/10885403512674, 1149-00/10885403517822, 1152-00/10885403517839, 1155-00/10885403521300, 1116-00/10885403512605
- Product quantity
- 53669
- Voluntary mandated
- Voluntary: Firm initiated
- City
- San Diego
- State
- CA
- Country
- United States
- Event id
- 97712