Paragon 28, Inc.: Phantom Fibula Nail, 2.8mm x 130mm, Right, Ti, Fibula Nail, REF: P36-128-130R-S
Read the official notice at FDA (device) →
The text below is copied from the official FDA (device) record. If anything differs, the official notice is correct.
- Company / brand
- Paragon 28, Inc.
- Product
- Phantom Fibula Nail, 2.8mm x 130mm, Right, Ti, Fibula Nail, REF: P36-128-130R-S
- Reason / hazard
- Fibula nail system intended for use in fixation of fibular fractures and osteotomies, has out of specification thread depths on the fibula nail, so implant may not be able to mate properly with the associated mounting bolt. If nail (implant) cannot properly mate with instrumentation, the surgery potentially will need to be aborted.
- Classification
- Class II
- Status
- Ongoing
- Distribution / sold at
- US Nationwide distribution in the states of NJ, WA, NY, MD, CA, FL, NM, TX, IL, OH, NV, TN, LA.
- Recall initiated
- 2025-10-13
- Code info
- UDI-DI: 00889795133670. Lot: 501182207125. Expiration: March 13th, 2030
- Product quantity
- 20
- Voluntary mandated
- Voluntary: Firm initiated
- City
- Englewood
- State
- CO
- Country
- United States
- Event id
- 97891